35
Eco/Classic Installation Guide and Technical Specification
7.3
Checking the Initial Fault Safety
To check the safety equipment, proceed as follows:
The switching positions I and II are testing settings used only to check the safety during
the annual inspection, or after repair work, or each time bed is put into service again.
• Setting switch position 4 (padlock symbol open ). Move all bed adjustments to a
slightly raised position.
• Setting switch position 3 (padlock symbol closed ). When operating the
adjustment buttons, no motorised adjustments should be possible.
• Set switch on the back of the handset to testing position 1 (symbol I).
• When operating the adjustment buttons, no motorised adjustments should be
possible.
• Set switch on the back of the handset to testing position 2 (symbol II).
• When operating the adjustment buttons, no motorised adjustments should be
possible.