32
Eco/Classic Installation Guide and Technical Specification
6.6
Qualification of Users
The care bed must only be operated by persons who have the corresponding training or
experience to enable them to handle the care bed correctly.
6.5
General Regulations
The care bed must only be used for the purpose intended. When setting up, operating,
and using the care bed, respect the regulations in your country, the general recognised
rules of technology, the occupational health and safety, and accident prevention
regulations.
If the care bed is in a faulty state causing the patient/occupant, care personnel or third
persons to be endangered, operation must not be started.
7. Servicing
7.1 Principles
Operators of care beds are obliged according to the regulations in your country.
The test according to the regulation EN 62353 contains the following minimum
requirements:
• Visual check
• Functional test
• Overall evaluation
To guarantee safe operation, a visual and functional test including an electrical test must
be carried out at least once a year. For this purpose, proceed according to the technical
safety checklist as per regulation EN 62353
IMPORTANT
If you have any doubts about the safety or functioning of the bed or even a part of the bed as
a result of the work performed below, the bed should under no circumstances be placed into
service again. Contact the supplier or manufacturer in this case.
7.2
List of Technical Safety Checks according to EN 62353
Care bed: Eco / Eco Low / Classic / Classic Low
Serial no: ………………………………
Location: ………………………….....
Person responsible: …………………………………………….
Inspected by: ……………………………………………..