Appendices
Appendix A – Product Specification
1. Display
Type:
Red/green, 7-segment, 5-digit primary display, 1-digit secondary display.
Height:
18mm (0.71inches) primary display, 7mm (0.3 inches) secondary display
Annunciators:
Alarm 1 and Alarm 2 status
2. Sensor
Dewpoint
Range:
From
–110°C
to
+20°C
Accuracy:
±
2°C over full range
Operating
Temperature:
Storage
Temperature:
Mechanical
Connections:
Electrical
Connection:
Male
coaxial
plug
Approvals:
CE,
Atex
-
3. Hygrometer Input
Sampling Rate:
10Hz
Resolution: 14-bits
Sensor short circuit detection:
Detected within two seconds. All alarms become active
4. Transistor Outputs
Type:
NPN open collector. Output 1 tied to Alarm 1. Output 2 tied to Alarm 2.
5. Relay Output 1
Contact Type:
Single pole, double throw
Rating:
2A
resistive,
120/240V,
50/60Hz
Lifetime:
>500,000 operations at rated voltage/current
Isolation:
Inherent
6. Relay Output 2
Contact Type:
Single pole, double throw
Rating:
2A
resistive,
120/240V,
50/60Hz
Lifetime:
>500,000 operations at rated voltage/current
Isolation:
Inherent
7. Linear (Re-Transmitted) Output
Ranges Available:
0-5V, 1-5V, 1-10V, 2-10V, 0-20mA and 4-20mA
Accuracy:
±
0.5% max of display value
Load
Impedance:
500
Ω
max for mA ranges, 500
Ω
min for voltage ranges
8. Serial Communications
Type:
Serial asynchronous, UART to UART
Data Format:
Open ASCII; one start bit, even parity, seven data bits and one stop bit.
Physical
Layer:
RS485
Maximum Number of zones
99
Baud Rate:
Selectable – 1200, 2400, 4800 or 9600
9. Performance
Reference Conditions
Ambient
Temperature: 20°C
±
2°C
Relative Humidity:
60 – 70%
Supply Voltage:
90 – 264V, 50Hz
Operating Conditions
Ambient
Temperature: 0°C
±
55°C
Relative Humidity:
20 – 95% non-condensing
10. Environmental
Approvals:
CE
EMC:
Certified
to
EN61326
Notes:
1) For RF fields (10V/m 80% Am, 1kHz) the reading accuracy may be impaired
by up to –0.3% in the frequency band 87 to 109MHz.
2) For line-conducted disturbances induced by RF fields (10V 80% AM 1kHz).
The instrument is self recoverable in the frequency band 0.15-0.73 Hz
Safety:
Complies
with
EN61010-1
Front Panel Seal:
To IP66
Supply Voltage:
90 – 264V, 50/60Hz
Power
Consumption:
7.5VA
max
11. Physical
Dimensions: Height
–
48mm
Width
–
95mm
Depth
–
110mm
without
sensor
connector.
Weight
0.21
kg
max.
Mounting:
Panel-mount; press-fit fixing. Panel cut out – 45mm x 92mm.
Terminals:
Screw
type
dewTEC
DS2000 User’s Manual
Page 22
Содержание dewTEC DS2000
Страница 2: ...dewTEC DS2000 User s Manual Page ii ...
Страница 4: ...dewTEC DS2000 User s Manual Page iv ...
Страница 6: ...dewTEC DS2000 User s Manual Page vi ...
Страница 29: ...Appendix B General Arrangement dewTEC DS2000 User s Manual Page 23 ...