5.3.2 Message
Form
2
This message implements a Parameter Read operation. The message from the master device is
of the form:
Laap?*
where
aa
is the address (a two digit hexadecimal number)
p
is a single character parameter identifier (see table below)
The reply, if the Parameter Read operation is successful, is of the form
LaapnnnnA*
where
aa
is the same address as in the received message
p
is a single character parameter identifier (see table below)
nnnnn
is the data (a five digit hexadecimal number)
If the specified parameter is invalid (e.g. not applicable to the addressed instrument), the reply
is of the form:
Laap00000A*
where
aa
is the same address as in the received message
p
is a single character parameter identifier (see table below)
Identifier Hex.
Parameter
Range
:
3A
Dewpoint
Depending on factory set range
;
3B
Total Value
-19999 to 99999
<
3C
Maximum Dewpoint
-19999 to 99999
=
3D
Minimum Dewpoint
-19999 to 99999
>
3E
Elapsed Alarm 1 Time
0 to 60000
E
45
Alarm 1 Dewpoint
Range Maximum to Range Minimum
F
46
Alarm 2 Dewpoint
Range Maximum to Range Minimum
5.4 Error
Conditions
If a slave detects a syntax error or parity error, it will not reply to the message; the master
device should make up to two retries, applying the two second No Reply timeout in each case.
Parameter read operations with parameter identifiers which are not applicable to the addressed
instrument, will have no effect and a positive acknowledgement will be returned.
dewTEC
DS2000 User’s Manual
Page 19