Error
Code
Description
Cause
E-
409
Sensor changed,
output turned off
The sensor type was changed, causing
the output to be turned off.
E-
410
Temperature was
out of tolerance,
output turned off
The temperature went out of tolerance
and the output was turned off.
E-
415
Sensor short,
output turned off
A sensor short circuit was detected
and the output was turned off.
E-
416
Cal failed
An error occurred during calibration
that caused the calibration to fail. The
unit is no longer in calibration mode.
E-
419
TEC not stable
The TEC is considered stable if the
temperature has changed less than
0.02°C for more than 20 seconds.
E-
433
Not a TEC
The TEC:CHAN command attempted to
select a non-TEC channel
E-
434
Ite limit exceeds
cable rating
The cable plugged into the unit is not
rated to carry the current as set by the
current limit.
E-
435
Mode Change
A mode change occurred when the
output was on, forcing the output off.
E-
436
AutoTune failed
The AutoTune process failed. See the
“Using the AutoTune Function” for
more information.
E-
437
AutoTune in T
mode only
The AutoTune process is only available
in T (temperature) mode. In R mode,
the PID parameters much be
determined manually.
E-
997
Control error, cycle
power
A hardware control error occurred,
cycle power to resolve. If error
continues to occur, contact factory.
E-
998
Command not
supported
A command was received that is not
supported by this instrument.
E-
999
Non-specific error
A non-specific error was encountered.