Page
14
· TC-3 Temperature Controller
User’s Manual
Menu
Description
Factory
Default
PID P
The proportional term of the PID loop.
Will only be available if
Gain
is set to
PID
.
1
PID I
The integral term of the PID loop. Will
only be available if
Gain
is set to
PID
.
0.01
PID D
The derivative term of the PID loop. Will
only be available if
Gain
is set to
PID
.
0
Sensor
Sets the temperature measurement
sensor. Can be
Therm 100uA
,
Therm
10uA
,
Therm,
LM335
,
AD590
,
RTD
, or
RTD (4-wire)
.
Therm
100uA
T-Low Lim
T-Low Lim
is the lower temperature
limit. If operating in T Mode and the
actual temperature drops below this
value, the output will be turned off.
Temperature limits not monitored in R
Mode or Ite Mode. Limit monitoring can
also be disabled in software (see
TEC:ENAB:OUTOFF command).
-99°C
T-High Lim
T-High Lim
is the upper temperature
limit. If operating in T Mode and the
actual temperature rises above this
value, the output will be turned off.
Temperature limits not monitored in R
Mode or Ite Mode. Limit monitoring can
also be disabled in software (see
TEC:ENAB:OUTOFF command).
125°C
R-Low Lim
R-Low Lim
is the sensor low limit. If
operating in R Mode and the actual
sensor measurement drops below this
value, the output will be turned off. R
limits are not monitored in Ite Mode.
Limit monitoring can also be disabled in
software (see TEC:ENAB:OUTOFF
command).
0.01kΩ