TC-3 Temperature Controller User’s Manual · Page
11
error, press
Menu
button. To clear all errors, press the knob. A list of
error codes can be found in the
Error Messages
section below.
When the unit is in remote mode, the yellow Remote LED will be lit.
More information about how the instrument behaves in remote mode
can be found in the Remote Mode Operation section below.
Control Modes
The
TC-3
offers three control modes: constant temperature mode (T
Mode), and constant resistance mode (R Mode), and constant current
mode (Ite Mode).
Changing the control mode is done through the menu by changing the
Mode
parameter to
T Mode
,
R Mode
, or
Ite Mode
.
Constant temperature mode uses the sensor constants to calculate the
resistance of the thermistor at the desired or actual temperature.
While most users will only need the temperature control mode, the
constant resistance mode allows you to bypass the sensor equation
and directly select the sensor set point. This can be useful when only
the desired sensor value is known, or when the sensor-to-temperature
conversion values are not available for your sensor.
Main Display Screen
On the main display you will find the set point and two or three
measurements, depending on the configuration you have chosen. An
example display is shown below:
Sample Display
The set point will depend on the control mode you have selected.
When in T Mode, the set point will be temperature. When in R Mode,
the set point will be in the active sensor’s units (ohms for thermistor
and RTD, uA for AD590, and mV for LM335). When in Ite Mode, the set
point will be in amps.
Measurements
The second line of the display contains the measurements appropriate
for the control mode. In T Mode, the actual temperature, current, and
voltage can be displayed, depending on the display mode. In R Mode,
the actual sensor value is displayed instead of the actual temperature.
In Ite Mode, the actual temperature is displayed.