35
OPERATION MANUAL
The average service life is 5 years.
The easily touched parts of the device are made of biosafety materials. Outer
surfaces of the device’s components are resistant to chemical disinfec on using any
disinfectant solu on approved for use in medical prac ce to protect the plas c and
metal items.
The maximum temperature on the surface of inductor coils contac ng with the
human body is not more than +41 °C, for the control unit surface, the maximum
temperature is +45 °C.
The exposure (magne c
fi
eld treatment dura on) is set automa cally to 20 min-
utes ±5% for all the treatment modes.
The sound signal indica ng the intermediate exposure me intervals will be gen-
erated every 5 minutes ±5% from the treatment start me and on.
The me delay before switching to standby mode a er the comple on of the
treatment is 5 minutes ±5%.
The device provides data storage in the internal non-vola le RAM including the
last set treatment mode saving and retrieval.
The le er «N» on the body of each inductor coil corresponds to the north pole
of the magne c
fi
eld induced by the inductor coils.
The hazard class dependent on the poten al applica on-associated risk is 2a
(moderate-rated risk medical product/device).
The device’s components’ dimensions and weight are given in Table 7.
Component name
Dimensions, mm
Weight, kg, not
more than
length
width
height
Control Unit
142±10
75±10
35±5
0.8
Emi er
890±15
88±5
18±5
Table 7