Bedienungsanleitung
Operation manual
Notice d’utilisation
Instrucciones de servcio
Istruzioni per l’uso
www.alluris.de
Page 21 of 34
4.4.4. Saving measuring cycles (Series)
Having set the Series feature, with start of the measurement cycle all
measured values are recorded.
During the recording the Mem symbol (and the unit symbol) in the
display flashes.
All measured values are saved according to the display-refresh rate. In
addition - if a peak mode is selected - the respective peak values are
captured in high-speed mode and stored. A series of measurements
can capture up to 1000 individual values.
You may pause the recording of measurement values while the
measuring keeps running and restart the recording, both by pressing
the S-key.
Recording ends with the termination of the measurement by pressing
the I-key or by the auto-stop function.
Restarting a measurement adds the newly captured values to the al-
ready stored values in the memory.
The memory content can be cleared completely by pressing for 2 se-
conds the O-key or by turning the instrument off.
Idle
Mem
Start
N
Peak Mem
pause
recording
measuring finished
last measured value
displayed
Stop
measurement
runs without
recording
Peak Mem
N
Peak Mem
N
N
continue
recording
measure and
record
Peak Mem
N
N
S
measure
and record
Peak Mem
N
N
S
Speichern von Messreihen
> <
measure and
record
Peak Mem
N
N
Start
etc.