3.5. Taring
On each start of the instrument the auto-tara function performs an
automatic taring to eliminate the gravitation force influence of the
weight of the sensor, the measuring axis and other installed attach-
ments according to the position of the instrument. At the same time
the temperature offset is readjusted.
The auto-tara-function can be deactivated, e.g. to continue a pausing
measurement
(see p. 18 chapter 4.2.7 Automatic taring function
. Due to the very high sensivity of the instrument
even slight movements or minute vibrations in the environment are
captured and displayed with values uneven zero.
In standard mode and in Peak mode you can tare the instrument du-
ring a measurement by pressing the O-key.
3.6. Displaying measured values
During measurements in standard mode the display shows the mea-
sured real-time value. In all of the three peak modes the instrument
captures and stores beside the actual measured value also the posi-
tive and negative peak value. In
Peak+
mode or
Peak –
mode either
the peak push value or peak pull value are displayed during measu-
rements according to the choosen prefix settings. In
Peak
mode the
real-time value is indicated during the measurement.
By pressing the S-key the display shows the measured values one
after the other. In
Peak+
mode first the positive peak value is shown.
By pressing the S-key next the negative peak value appears and after
pressing the S-key once again the last measured value is displayed.
In
Peak –
the display sequence starts with the negative peak value. If
the measurement has been taken in
Peak
mode (series FMI-B20 and
FMI-S20 and higher), the Peak symbol flashes accordingly.
You start the next measurement without deleting the peak memory, by
pressing the I-key. Peak values remain stored in the memory as long
as a higher peak value is measured.
S
peak value
negative peak
value
last
value
(FMI-x20 and higher)
Peak
cN
Peak
cN
Peak
cN
S
Peak modes:
Display sequence of stored values