1. Working Conditions
Environmental Temperature: 10
C
~
40
C
Relative Humidity: ≤80%
Input Voltage: AC 100
~
240V, 50Hz/60Hz
2.
Basic Parameters
Table 1 Parameters
Model
Parameters
Auto-Pure Mini
Principle
Magnetic Particle Method, Magnet type
Sample Volume
50μL—1000μL
Throughput
1—16
Stability
CV≤3%
Extraction time
With different nucleic acid extraction reagents, 10 ~ 60min/time
Temperature control
module
Ambient temperature ~ 120
℃
for lysis and elution
Heating time
Ambient temperature ~120
℃
≤5 minutes
Temp. Accuracy
±1
℃
Operation
4.3 inch touch screen
APP Software
Equipped with Android APP software to realize protocol editing
and data transmission with the instrument
Internal protocols
Store up to 100 groups of protocols
Extension interface
With USB port, ethernet port and RS232 port
Data transfer
Support Bluetooth, USB, RS232 data transmission, can expand the
transmission of PC port
Bar code scan
External code scanning gun, software can expand the sample
information input