![Allsheng Auto-Pure Mini Скачать руководство пользователя страница 30](http://html1.mh-extra.com/html/allsheng/auto-pure-mini/auto-pure-mini_operation-manual_2903938030.webp)
Auto-Pure Mini Operation Manual
Chapter 4 Operations
24
Fig 31 Option settings
The setting meanings of each option are shown in the following table;
Table 5 Option settings
Step
Option
Meaning
Heating
settings
Heating in advance
When running to the heating hole position, the
magnetic rod's tip or magnetic rod will start to action
after the temperature of the heating hole reaches the
preset temperature
Heating and action
synchronized
When running to the heating hole position, heating is
carried out synchronously with the magnetic rod's tip
or magnetic rod action
Action after reach
specified T
When running to the heating hole position, the
magnetic rod's tip or magnetic rod start action after
reach the set temperature
Elution well heated
when the protocol
run to the setted
step
Set advanced steps to heat, up to 3 steps
Cooling
settings
Cooling in advance
When running to this step, if the actual temperature is
higher than the set temperature, the temperature will
be cooled first, and then the magnetic rod's tip or
magnetic rod action will be started
Cooling and action
synchronized
When running to this step, cooling is carried out
synchronously with the magnetic rod's tip or magnetic