Allpro ScanX trek D5000-T Скачать руководство пользователя страница 4

ALLPRO Imaging

Page 4

ScanX has been designed to minimize exposure of personnel to hazards . While the equipment is designed 
for safe operation, certain precautions must be observed . Use of the ScanX not in conformance with the 
instructions specified in this manual may result in permanent failure of the unit .

General.

  Check with your dealer for packing material requirements if it is necessary to return the product 

to the manufacturer . Correct packing guarantees optimal safety of the device during transport . 
Should it become necessary to return the device to the manufacturer during the warranty period, 
the manufacturer will not accept claims for damage arising from using incorrect packing materials . 

  Before every use, the operator must check the functional safety and the condition of the device .

  The operator must be knowledgeable in the operation of the device . 

Use of Accessory Equipment.

The use of ACCESSORY equipment not complying with the equivalent safety requirements of this 
equipment may lead to a reduced level of safety of the resulting system . 
Use of ACCESSORIES or cables other than those specified or provided by the manufacturer may result 
in increased EMISSIONS or decreased IMMUNITY of the EQUIPMENT .

Do Not Attempt Internal Service.

The interior of each component of ScanX is only accessible by removing hardware with tools and 
should only be opened and serviced by an authorized dealer service technician .
Contact your local authorized dealer for service . Failure to heed this directive may result in equipment damage 
and voiding the warranty .

Electrical Safety Notes.

  The main power switch is the main power disconnect device .

  Use only the line cord provided with the unit .

  Use only grounded electrical connections .

  To avoid risk of electric shock, fire, short-circuit or dangerous emissions, never insert any metallic 

object into the equipment .

  Only use connection cable(s) delivered with the device .

  Check the device cables for possible damage before switching on . Damaged cables, plugs and 

sockets must be replaced before use .

  Never touch open supply outlets and patients simultaneously .

  Do not locate unit where it could be sprayed with water, or in a damp environment .

GENERAL SAFETY

Содержание ScanX trek D5000-T

Страница 1: ...Operator s Manual Di g i tal Ima ging Sy s te m Part Numbers D5000 T and D5000 TB ScanX Trek without Battery D5000 T ScanX Trek with Battery D5000 TB...

Страница 2: ......

Страница 3: ...y listed below including devices with or without battery power Device Name Part No No Battery With Battery Trek D5000 T D5000 TB Designed and built exclusively for the demanding conditions encountered...

Страница 4: ...y lead to a reduced level of safety of the resulting system Use of ACCESSORIES or cables other than those specified or provided by the manufacturer may result in increased EMISSIONS or decreased IMMUN...

Страница 5: ...in hazardous radiation exposure The laser is on only during an active scan Only a trained technician from an authorized dealer should remove a cover from the ScanX Direct eye contact with the output...

Страница 6: ...RE TO BEAM CLASS IIIb LASER PRODUCT COVER REMOVED MAKES THIS DEVICE A The ScanX is a Class I Laser Product Class 1 Laser Product IEC This warning label identifies the ScanX as such a product and descr...

Страница 7: ...h regard to safety reliability and proper functioning only if assembly resetting changes or modifications and repairs have been carried out by an authorized dealer and if the equipment is used in conf...

Страница 8: ...e received If any item is missing notify your dealer UNPACKING AND INSPECTION Figure 1 Typical ScanX Trek Product Family Packaging ScanX Unit Accessory Kit Accessory Kit Contents Accessory Literature...

Страница 9: ...or for complete mobility inserted into either an optional hard or soft carrying case for easy and safe transport anywhere image scanning is required Rechargeable Battery The ScanX Trek has a model wi...

Страница 10: ...midity 5 95 Non condensing Note Resolution of the ScanX is dependent on operating mode and specific imaging plate type used Resolution LP mm Horizontal Vertical 3 8 to 5 0 3 8 to 7 Compliance Data Las...

Страница 11: ...x64 processor Microsoft Windows 7 Professional Enterprise or Ultimate with Service Pack 1 for an Intel 32 bit or an Intel 64 bit extended x64 processor Microsoft Windows 8 1 Professional or Enterprise...

Страница 12: ...computer cm centimeter GB gigabyte 230 109 bytes GHz Gigahertz 109 of Hertz H height Hz Hertz cycles per second IEC International Electro technical Commission IMS Image Management Software IP imaging...

Страница 13: ...t in the ON position depressed and flush with the panel this switch allows the ScanX to operate from battery power While engaged the CHARGE LEVEL Indicator illuminates 2 When placed in the OFF positio...

Страница 14: ...ote When the Eraser Status Indicator displays a burst of approximately 5 quick flashes it is alerting the user that the erase function is reduced due to a high temperature condition in the red LED era...

Страница 15: ...11 necessary to support ScanX operation Set up the computer according to the manufacturer s recommendations 2 Verify that an authorized Imaging Software and the supplied USB drivers are installed prop...

Страница 16: ...e internal battery must always be charged prior to using To reach the full operational power level charge for at least 3 hours before using for the first time and on subsequent uses Refer to figure 6...

Страница 17: ...with the plates CAUTION Use a Plate Protector for each plate PSPs may be exposed in cassettes and scanned without Plate Protectors however greater care must be exercised to not expose the PSP to light...

Страница 18: ...ll be acquired Method 2 Perform the Activate Scanner and the Scanning and Erasing Plates procedures on pages 19 and 20 This method scans the plate and the imaging software may acquire a junk image sca...

Страница 19: ...the ScanX The PSP is now ready to be scanned Activate Scanner Refer to Figure 9 and activate the ScanX via AC power or the internal Battery power by performing the fol lowing procedures 1 Refer to Fi...

Страница 20: ...by both guide flaps resting on the inlet ring which further aids in alignment 5 At this point all four track lights will turn yellow indicating the PSP has been sensed and the Scanner is transporting...

Страница 21: ...indicate that the Erase function is OFF The PSP will not be erased after scanning 4 Insert the PSP to be scanned into the ScanX by performing the Scanning and Erasing Plates procedures provided on the...

Страница 22: ...not run Verify Computer System requirements Page 11 Make sure that the READY switch is set to ON and the green indicator is lit Reconnect the cable Check for tightness Replace if necessary Verify tha...

Страница 23: ...long in cassette Partial image erasure due to exposure to light during han dling of the PSP Make sure the ScanX is operating with both eraser strips turned on blue LED indicator below the ERASER switc...

Страница 24: ...no effect on the quality of the image and should be expected under normal conditions If the phosphor side is scratched make sure the plates are being handled properly and not being dragged from the Sc...

Страница 25: ...pense the malfunctioning system back to ALLPRO Imaging in the same shipping container All equipment under warranty require a Return Materials Authorization RMA issued by Technical Services Items retur...

Страница 26: ...ALLPRO Imaging Page 26 NOTES...

Страница 27: ...Page 27 ALLPRO Imaging NOTES...

Страница 28: ...and software to facilitate smooth and efficient radiography procedures Thousands of ALLPRO Imaging products have been sold globally installed and serviced through our network of authorized dealers Pr...

Отзывы: