Allpro ScanX trek D5000-T Скачать руководство пользователя страница 18

ALLPRO Imaging

Page 18

IMPORTANT:

 PSPs must always be erased prior to use.

Note:

  Use PSPs within 24 hours of last erasure. Repeat erasing process if PSPs have been stored longer than 

24 hours. 

Erasing PSPs

Each PSP should be used (i .e . X-ray exposed and scanned) 

within 24 hours

 of erasure since natural 

radiation will add noise to the PSP . Erase PSPs by simply using the ScanX In-Line Erase Feature . Erasing 
of PSPs can be accomplished using one of two methods as follows: 

Note:  Both erasing methods will result in an erased PSP suitable for reuse. The user will not observe any 

difference in ScanX operation when using either method.

Method #1 
 

Perform the Activate Scanner and the Scanning and Erasing Plates procedures on pages 19 and 
20 .  Except when performing step 4 of the Activate Scanner procedure, select the Erase option 
from the installed authorized imaging software to activate the ScanX . This method does not scan 
the plate and no image will be acquired .

Method #2 
 

Perform the Activate Scanner and the Scanning and Erasing Plates procedures on pages 19 
and 20 .  This method scans the plate and the imaging software may acquire a “junk image” 
(scanned latent plate image) that should be subsequently deleted from the imaging software .

Cleaning Phosphor Storage Plates

For the best images, PSPs and Plate Protectors should be handled carefully and kept clean . Use specially 
formulated PSP Cleaning Wipes (P/N B8910) to clean all PSPs and  Plate Protectors . These single-use 
extra soft, 100% polyester fabric wipes will not scratch or damage while safely removing dust, hair, dirt 
and smudges from the imaging surface . Use one wipe and clean plates as follows:

1 .  As shown by Figure 8, gently wipe the PSP Cleaning 

Wipe over the dry Plate surface . Wipe back and forth 
and then in a circular motion .

2 .  Allow the plate surface to air dry . Make sure that the 

PSP is completely dry before re-using .

Disposal of Phosphor Storage Plates

Consult with your federal, national, state and local government, 
for rules and regulations on disposal of Phosphor Storage Plates .

Figure 8 . PSP Cleaning

PLATE CARE & PREPARATION

Содержание ScanX trek D5000-T

Страница 1: ...Operator s Manual Di g i tal Ima ging Sy s te m Part Numbers D5000 T and D5000 TB ScanX Trek without Battery D5000 T ScanX Trek with Battery D5000 TB...

Страница 2: ......

Страница 3: ...y listed below including devices with or without battery power Device Name Part No No Battery With Battery Trek D5000 T D5000 TB Designed and built exclusively for the demanding conditions encountered...

Страница 4: ...y lead to a reduced level of safety of the resulting system Use of ACCESSORIES or cables other than those specified or provided by the manufacturer may result in increased EMISSIONS or decreased IMMUN...

Страница 5: ...in hazardous radiation exposure The laser is on only during an active scan Only a trained technician from an authorized dealer should remove a cover from the ScanX Direct eye contact with the output...

Страница 6: ...RE TO BEAM CLASS IIIb LASER PRODUCT COVER REMOVED MAKES THIS DEVICE A The ScanX is a Class I Laser Product Class 1 Laser Product IEC This warning label identifies the ScanX as such a product and descr...

Страница 7: ...h regard to safety reliability and proper functioning only if assembly resetting changes or modifications and repairs have been carried out by an authorized dealer and if the equipment is used in conf...

Страница 8: ...e received If any item is missing notify your dealer UNPACKING AND INSPECTION Figure 1 Typical ScanX Trek Product Family Packaging ScanX Unit Accessory Kit Accessory Kit Contents Accessory Literature...

Страница 9: ...or for complete mobility inserted into either an optional hard or soft carrying case for easy and safe transport anywhere image scanning is required Rechargeable Battery The ScanX Trek has a model wi...

Страница 10: ...midity 5 95 Non condensing Note Resolution of the ScanX is dependent on operating mode and specific imaging plate type used Resolution LP mm Horizontal Vertical 3 8 to 5 0 3 8 to 7 Compliance Data Las...

Страница 11: ...x64 processor Microsoft Windows 7 Professional Enterprise or Ultimate with Service Pack 1 for an Intel 32 bit or an Intel 64 bit extended x64 processor Microsoft Windows 8 1 Professional or Enterprise...

Страница 12: ...computer cm centimeter GB gigabyte 230 109 bytes GHz Gigahertz 109 of Hertz H height Hz Hertz cycles per second IEC International Electro technical Commission IMS Image Management Software IP imaging...

Страница 13: ...t in the ON position depressed and flush with the panel this switch allows the ScanX to operate from battery power While engaged the CHARGE LEVEL Indicator illuminates 2 When placed in the OFF positio...

Страница 14: ...ote When the Eraser Status Indicator displays a burst of approximately 5 quick flashes it is alerting the user that the erase function is reduced due to a high temperature condition in the red LED era...

Страница 15: ...11 necessary to support ScanX operation Set up the computer according to the manufacturer s recommendations 2 Verify that an authorized Imaging Software and the supplied USB drivers are installed prop...

Страница 16: ...e internal battery must always be charged prior to using To reach the full operational power level charge for at least 3 hours before using for the first time and on subsequent uses Refer to figure 6...

Страница 17: ...with the plates CAUTION Use a Plate Protector for each plate PSPs may be exposed in cassettes and scanned without Plate Protectors however greater care must be exercised to not expose the PSP to light...

Страница 18: ...ll be acquired Method 2 Perform the Activate Scanner and the Scanning and Erasing Plates procedures on pages 19 and 20 This method scans the plate and the imaging software may acquire a junk image sca...

Страница 19: ...the ScanX The PSP is now ready to be scanned Activate Scanner Refer to Figure 9 and activate the ScanX via AC power or the internal Battery power by performing the fol lowing procedures 1 Refer to Fi...

Страница 20: ...by both guide flaps resting on the inlet ring which further aids in alignment 5 At this point all four track lights will turn yellow indicating the PSP has been sensed and the Scanner is transporting...

Страница 21: ...indicate that the Erase function is OFF The PSP will not be erased after scanning 4 Insert the PSP to be scanned into the ScanX by performing the Scanning and Erasing Plates procedures provided on the...

Страница 22: ...not run Verify Computer System requirements Page 11 Make sure that the READY switch is set to ON and the green indicator is lit Reconnect the cable Check for tightness Replace if necessary Verify tha...

Страница 23: ...long in cassette Partial image erasure due to exposure to light during han dling of the PSP Make sure the ScanX is operating with both eraser strips turned on blue LED indicator below the ERASER switc...

Страница 24: ...no effect on the quality of the image and should be expected under normal conditions If the phosphor side is scratched make sure the plates are being handled properly and not being dragged from the Sc...

Страница 25: ...pense the malfunctioning system back to ALLPRO Imaging in the same shipping container All equipment under warranty require a Return Materials Authorization RMA issued by Technical Services Items retur...

Страница 26: ...ALLPRO Imaging Page 26 NOTES...

Страница 27: ...Page 27 ALLPRO Imaging NOTES...

Страница 28: ...and software to facilitate smooth and efficient radiography procedures Thousands of ALLPRO Imaging products have been sold globally installed and serviced through our network of authorized dealers Pr...

Отзывы: