ALL-TEST
PRO
®
31 User Manual
©2011, ALL-TEST
Pro,
LLC
www.alltestpro.com
rev 2011-11F
36
Capacitor Testing
Note: This is a Pass/Fail test.
Motor starting capacitors and power factor correction capacitors can be effectively tested with the ALL-
TEST
PRO®
31. This test will not provide the capacitance value of the test specimen, but will indicate
if the capacitor is performing properly in an AC Circuit.
1.
Turn line power
off
. Disconnect one capacitor lead from the circuit.
2.
Short the capacitor terminals to remove any remaining charge.
3.
Put the ALL-TEST
PRO®
31 in “winding test” mode by selecting the Motor icon from the mode
screen.
4.
Connect the red and black test leads to the capacitor.
5.
An impedance value should be displayed. If impedance value shows “---”, the capacitor is open.
6.
Observe the phase angle measurement in the upper right of the display. If the capacitor is good,
this number is very close to 90.
7.
Press the TEST key. For a good capacitor, the main display will show 90 or above. A reading
below 90 indicates a poor or failed capacitor.
8.
Turn the instrument off.