44
Serial Communications
Permanently erase one or more stored measurements by typing:
Erase measurement:
[unit ID]
me
[measurement ID]
<CR>
Example:
ame 167<CR>
(erases stored measurement #167)
Erase all:
[unit ID]
me<CR>
Example:
ame<CR>
(erases all stored measurements)
Quick Serial Command Guide
Note: Serial commands are not case-sensitive. For simplicity, we
assume that the unit ID of the FP-25 is A in the listing that follows.
Change the unit ID:
[current unit ID]
@=
[desired unit ID]
<CR>
Set clock time and date:
amcs
[sec.] [min.] [hour] [day] [month] [year]
<CR>
Tare volumetric flow:
av<CR>
Tare abs. pressure with baro:
apc<CR>
Poll the live data frame:
a<CR>
Begin streaming data:
[unit ID]
@=@<CR>
or
*@=@<CR>
Stop streaming data:
@@=
[desired unit ID]
<CR>
Set streaming interval:
aw91=
[number of milliseconds]
<CR>
Begin measurement:
amb
[duration in seconds]
<CR>
Abort current measurement:
ama<CR>
Display current/last meas.:
amd<CR>
Display measurement by ID:
amd
[measurement ID]
<CR>
List all stored measurement IDs:
ams<CR>
Delete all stored measurements:
ame<CR>
Delete measurement by ID:
ame
[measurement ID]
<CR>
Query manufacturer info:
a??m*<CR>
Query live data info:
a??d*<CR>
Query measurement info:
a??dm*<CR>
Lock the front display:
al<CR>
Unlock the front display:
au<CR>
Additional Serial Commands
If you have need of more advanced serial communication commands, please
contact Alicat.