34
Navigating and Customizing Your FP-25
Choosing Device Engineering Units from the Basic Configuration Menu
Changing device engineering units alters both the display and the data
frame. First choose the parameter whose unit you want to change, and
then choose your desired engineering unit. Confirm the change on the last
screen. Changing
engineering units
in the Pressure unit
menu changes
the units for both
absolute and
gauge pressure
readings, but
not barometric
pressure.
Setup |
Basic Configuration
MAIN
AVG
TIME
BACK
STP/
NTP
DEVICE
UNITS
MAIN Returns to the Instant
screen of Standard Mode or
Basic Mode, whichever was
last used.
BACK Returns to the top-level
Menu (page 31).
AVG TIME Sets the duration
for averaged measurements,
up to 9h:59m. Use the SELECT
DIGIT and UP/DOWN buttons to
choose the length of the timer
countdown. Press SET to accept
the new value. The time format
is chosen in MENU > BASIC CONFIG
> DEVICE UNITS > Time.
Basic Configuration Menu
The Basic Configuration Menu contains options for choosing the duration and
time format for the averaging function, as well as device engineering units.
DEVICE UNITS Changes device engineering
units for any parameter:
Standard Mass Flow - Volumetric Flow - Pressure -
Baro Pressure - Temperature - Time
SELECT
BACK
Pressure
DOWN
UP
Baro Pressure
Temperature
Average Time
Volumetric Flow
Standard Mass Flow
>
SET
CANCEL
PRESSING SET WILL
AFFECT DISPLAY
AND SERIAL VALUES.
VERIFY CONNECTED
SERIAL DEVICES
EXPECT THE CHANGE.
0:00:05
Averaging time: h:m:s
SET
CLEAR
BACK/
CANCEL
SELECT
DIGIT
DOWN
UP
>
STP/NTP Defines standard
(STP)and normal (NTP)
temperature and pressure