Navigating and Customizing Your FP-25
Averaged Data in Standard Mode
The averaging screen of Standard Mode displays averaged data for all flow
parameters simultaneously. Flow data is measured at a rate of 1000 readings
for every second of measurement time. Press the button above or below any
of the four flow parameters to highlight its value in the center of the screen,
even after an averaged measurement has finished. Engineering units cannot
be changed from the averaging screen.
• When an average is running, the screen displays Running Avg at the top, and
all parameter values update in real time until the average has completed.
When the countdown timer stops, the screen displays Final Avg at the top.
Note: If you have chosen the time remaining to be highlighted in the
center of the display, this parameter will disappear once the averaged
measurement is finished, just as its smaller counterpart in the top right
of the display disappears and becomes
START AVG
. The time remaining
parameter will reappear in the center when you start a new average.
Standard Mode
| Averaged Data
Temp Avg
+24.38
#C
RH 38%
LPM
Probe
Vol Flow Avg
+16.667
Final Avg
CANCEL
AVG
+15.442
SLPM
+16.667
LPM
REMAIN
0:03
#C
+24.38
BARO
+704.00
Highlights remaining time in
the center:
CANCEL AVG Stops the current
averaged measurement.
Returns to Instant screen.
Std Flow Avg
+15.442
SLPM
RH 38%
Highlights standard (mass)
flow average in the center:
Vol Flow Avg
+16.667
LPM
RH 38%
Highlights volumetric (actual)
flow average in the center:
Highlights temperature
average in the center:
Baro Press Avg
+704.0
mmHgA
RH 38%
Gage Press Avg
-0.12
PSIG
RH 38%
Abs Press Avg
+13.50
PSIA
RH 38%
Highlights pressure
average in the center:
Internal absolute pressure:
Internal gauge pressure:
Barometric pressure:
Time Remain
0:03
h:m:s
RH 38%
INST/MENU Returns to Instant
screen. Appears after a
measurement has completed.
START AVG Begins another
averaged measurement.
Appears after a measurement
has completed.
Averaging status display:
Running Avg = measuring a
single average
Final Avg = completed