3.11 Points Reviseion 

 
 
 
 
 
 
 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Speed_1  m/s 
 
                   0

0

.100    

 

Revise_1 
 
                   

1

.00000  

 

Speed_2  m/s 
 
                     

0

0.500  

 

Revise_2  
 

 

                   

1

.00000    

 

Speed_3  m/s 

 

 

 

                     

0

1.000    

 

Revise_3  
 

 

                   

1

.00000    

 

Speed_4  m/s 

 

 

 

                     

0

2.000    

 

Revise_4  
 

 

                   

1

.00000    

 

Speed_5  m/s 

 

 

 

                     

0

4.000   

 

Revise

 

Flow accuracy curve correctFactor,11 points 
Speed_1 ~ Speed_11  presents the flow velocity, you can set 11 point single flow velocity 
Revise_1 ~ Revise_11 separately to corresponding  Speed_1 ~ Speed_11, it use on amend the 
inaccuracy by  the flow velocity amend. 
Use 

 button to move, 

 button to change the value, Any related setting sample please see 

 below. 
ps: normally flowmeter accuracy it’s during  standard accuracy, only if it’s necessary, please do  
not adjust any accuracy curve or you can ask Alia professional engineer to adjust with you 

 

Example 1

: size :50 mm(2”), Calibration flowrate are:0.5 M3/Hr, 1 M3/hr, 2 M3/Hr, 4 

M3/hr,  
totally 4 points, first calibration result as follow: 
 
 
 
 
 
 
 
 
 
 
 
 
 
4point flowrate, its inaccuracy calculation as follow: 
   Point 1: 0.5/ 0.530 = 0.943 , new Revise valve = 0.943 
   Point 2: 1 / 0.983   = 1.017,  new Revise valve= 1.017 
   Point 3: 2 / 2.046   = 0.978,  new Revise valve= 0.978 
   Point 4: 4 / 4.176   = 0.958,  new Revise valve= 0.958 
 
Filled the 4 points Revise valve in order to the new flow Velocity adjust parameter as 
follow: 
 
 
 
 
 
 
You will get the digit  : 
Speed_1= 0.071, Revise_1=0.943, Speed_2=0.142,  Revise _2=1.017 
Speed_3= 0.283, Revise_3=0.978, Speed_2=0.566,  Revise _4 =0.958 
 
 
Separately filled the converter related parameter,then the adjustment finished, that  
times Revise_5, Revise _6, Revise _7 , Revise _8 , Revise_9, Revise _10,  
Revise _11, Speed_5,  speed_6 , Speed_7,  speed_8 , Speed_9,  speed_10   
speed_11  doesn’t need to change 
 
 
 
 

 
 

 

 

Measure 

point 1

 

Measure 

point 2

 

Measure 

point 3

 

Measure 

point 4

 

 

Actual 

flowrate

 

0.5 m

3

/h 1 

m

3

/h

 

2 m

3

/h

 

4 m

3

/h

 

 

Actual 

Velocity

 

0.071 m/s

0.142 m/s

0.283 m/s

0.566 m/s

 

 

Flowmeter 

flowrate

 

0.530 m

3

/h

0.983 m

3

/h

2.046 m

3

/h

4.176 m

3

/h 

 

 

Flowmeter 

Velocity

 

0.075 m/s

0.139 m/s

0.289 m/s

0.591 m/s

 

 

 

1

2

3

4

 

 

Speed

 

0.071

 

0.142

 

0.283

 

0.566

 

 

     Revise

 

0.943 1.017

 

0.978

 

0.958

 

ENT

 

Revise 
 
    NO        

YES 

          

 

Revise 

Flow accuracy curve correct Factor,11 points  
Choose  

YES 

 and shift to 

11points revision.

 

ENT

 

ENT

 

ENT

 

ENT

 

ENT

 

ENT

 

ENT

 

ENT

 

ENT

 

Содержание AMC2100 Series

Страница 1: ...ALIA TECHNOLOGY LLC...

Страница 2: ...verter AMC2100 Operate Manual 7 3 1 Converter AMC2100 Operate Manual 7 3 1 1 Power and signal output connected line compact and separate model 7 3 1 2 Separate Wiring 8 3 1 3 AMC2100 Panel deploy 8 3...

Страница 3: ...3 5 15 95 155 3 5 20 105 160 4 5 25 115 166 4 5 32 140 180 6 5 40 150 190 7 0 50 165 201 9 5 65 185 220 12 80 200 235 15 100 220 254 17 125 250 284 21 150 300 285 314 28 200 350 340 369 36 250 400 395...

Страница 4: ...tion Wafer 1 3 AMF301 Installation 1 4 AMF500 Installation mm Inch L D 25 1 90 71 32 1 1 4 80 40 1 1 2 86 50 2 115 100 65 2 1 2 115 120 80 3 130 131 100 4 155 151 125 5 155 181 150 6 185 206 200 8 215...

Страница 5: ...nge should be flat having no sharp residue Otherwise liner will be damaged After flowmeter is installed if other places in pipe needs to be welded flowmeter s power must be shut down 2 Usually there w...

Страница 6: ...and downstream of the installation should satisfied the following conditions Picture Generally tube installation the flowmeter Upstream could try to stay 5D 10D Downstream 2D 5D 2 4 Install method a...

Страница 7: ...he flowmeter tube it may cause Liner inside broken Please DO NOT use your hand or rope to hang on Converter or wiring box if Flowmeter size is bigger than 80mm because the material of converter and wi...

Страница 8: ...cide if you need to connect according to your needs 1 20mA output 4 2 4 20m output A 3 Pulse output 4 ulse P output 5 RS485 output 6 RS485 output Cable from sensor Power supply 90 260V AC 50 60Hz or 2...

Страница 9: ...level Save presently settled parameter Jump to next parameter setting Up Selected one of 4 display line Change present digit dot and survey unit Up Down movement Right Change the selected line on the...

Страница 10: ...u need Line 1 display from m3 h turn to 3 4 Totalizer Reset 123 456 m3 0 0000m3 h 123 456 m3 0 0000m3 h In normal display press button until Line 2 is selected 123 456 m3 0 0000 FS Press ENT button to...

Страница 11: ...f message window Power off message window ON date time when power on OFF date time when power off Q the flowrate when power off the totalizer when power off Example 01 means the power off recored seri...

Страница 12: ...mping s Hydro State Baudrate Data Bit Parity Stop Bit Modbus Address 0000 Direction MODBUS Type 123 456 m3 0 0000m3 h OPERATIONS SET ZERO ESC Empty Detect Empty Limit Communication ENT ENT ENT ENT ENT...

Страница 13: ...nsor Factor ENT ENT ENT ENT ENT Convertor Factor Coil Current A Restore Shipping Convertor NO ENT ENT ENT Loop Test Pulse Test Coil Test Password Coil Freq Hz Zero Factor ENT ENT ENT ENT 4mA Trim ENT...

Страница 14: ...ress or button to selected the output 1 Default 2 Pulse 3 Frequency 4 Alarm Pulse output Each pulse match the Liter value Liter Pulse Press button to move Press to changed the Value Pulse width Press...

Страница 15: ...if it needs empty pipe detect Empty pipe detect the value Press button to move Press to change the value Max Limit fs 05 Response speed adjustment Press button to move Press button to change the value...

Страница 16: ...Press button to move Press button to change the digit TIME hh mm ss 17 09 49 Time Setting Press button to move Press button to change the digit Pls press ENT button for 5 seconds then it will go back...

Страница 17: ...ART Address 00 DATE yy mm dd 08 01 10 Date Setting Press button to move Press button to change the digit ENT TIME hh mm ss 17 09 49 Time Setting Press button to move Press button to change the digit P...

Страница 18: ...tton to select the simulated output value 4 8 12 16 20 Password 1111 Password login window Input the system password Password 1111 OPERATIONS SET ZERO ESC ENT Pulse Simulated output Press or button to...

Страница 19: ...ose NO to skip to Restore Factory Settings choose YES to conduct Full Tube Trim If there is full tube empirical value you can choose Keyin to modify full tube frequency value Otherwise it is not recom...

Страница 20: ...ess or button to select the calibration type 4 8 12 16 20 Pulse Simulated output Press or button to select the calibration type 1 10 100 1000 2000 5000 Excit voltage calibration Check the coil X to Y...

Страница 21: ...ess button to move Press to change the value Pls press ENT button for 5 seconds then it will go back to the normal display Restore the parameter to second memory Press or button to select if it needs...

Страница 22: ...Revise valve 0 943 Point 2 1 0 983 1 017 new Revise valve 1 017 Point 3 2 2 046 0 978 new Revise valve 0 978 Point 4 4 4 176 0 958 new Revise valve 0 958 Filled the 4 points Revise valve in order to...

Страница 23: ...m3 h the actual speed was 5 66 m s display 4012 m3 h Velocity 5 677 m s New Revise 4000 4012 0 997 Flow Velocity 5 66 m s during Revise_5 4 m s and Revise_6 6 m s so you can set as follow Speed_5 5 66...

Страница 24: ...lected SET Alarm Select MAX MIN BAT DIR Alarm Select Press or button to BAT 1 Max Upper Alarm 2 Min Lower Alarm 3 Batch 4 Direction ENT Low Cutoff fs 0 0 Function for cutoff the low flowrate Press but...

Страница 25: ...on to AUTO or MANUAL ENT SET 1 000m3 RUN 0 000m3 START CLR EXIT Press or button to START To start batch control If choose CLR The actual flow value will be zero ENT SET 1 000m3 RUN 0 000m3 PAUSE In th...

Страница 26: ...lue of alarm setting Increase range Turn off alarm output or decrease the alarm lower limit value 256 EMPTY Empty pipe alarm Check the pipe full the pipe 512 Zero Zero mv value 99mv Re zero when fluid...

Страница 27: ...26 633 W 5th Street 26th Floor Los Angeles CA 90071 USA Tel 1 213 533 4139 Fax 1 213 223 2317 URL www alia inc com Email alia inc com alia...

Отзывы: