![Alia AMC2100 Series Скачать руководство пользователя страница 22](http://html1.mh-extra.com/html/alia/amc2100-series/amc2100-series_operation-manual_2896251022.webp)
21
3.11 Points Reviseion
Speed_1 m/s
0
0
.100
Revise_1
1
.00000
Speed_2 m/s
0
0.500
Revise_2
1
.00000
Speed_3 m/s
0
1.000
Revise_3
1
.00000
Speed_4 m/s
0
2.000
Revise_4
1
.00000
Speed_5 m/s
0
4.000
Revise
Flow accuracy curve correctFactor,11 points
Speed_1 ~ Speed_11 presents the flow velocity, you can set 11 point single flow velocity
Revise_1 ~ Revise_11 separately to corresponding Speed_1 ~ Speed_11, it use on amend the
inaccuracy by the flow velocity amend.
Use
button to move,
button to change the value, Any related setting sample please see
below.
ps: normally flowmeter accuracy it’s during standard accuracy, only if it’s necessary, please do
not adjust any accuracy curve or you can ask Alia professional engineer to adjust with you
Example 1
: size :50 mm(2”), Calibration flowrate are:0.5 M3/Hr, 1 M3/hr, 2 M3/Hr, 4
M3/hr,
totally 4 points, first calibration result as follow:
4point flowrate, its inaccuracy calculation as follow:
Point 1: 0.5/ 0.530 = 0.943 , new Revise valve = 0.943
Point 2: 1 / 0.983 = 1.017, new Revise valve= 1.017
Point 3: 2 / 2.046 = 0.978, new Revise valve= 0.978
Point 4: 4 / 4.176 = 0.958, new Revise valve= 0.958
Filled the 4 points Revise valve in order to the new flow Velocity adjust parameter as
follow:
You will get the digit :
Speed_1= 0.071, Revise_1=0.943, Speed_2=0.142, Revise _2=1.017
Speed_3= 0.283, Revise_3=0.978, Speed_2=0.566, Revise _4 =0.958
Separately filled the converter related parameter,then the adjustment finished, that
times Revise_5, Revise _6, Revise _7 , Revise _8 , Revise_9, Revise _10,
Revise _11, Speed_5, speed_6 , Speed_7, speed_8 , Speed_9, speed_10
speed_11 doesn’t need to change
Measure
point 1
Measure
point 2
Measure
point 3
Measure
point 4
Actual
flowrate
0.5 m
3
/h 1
m
3
/h
2 m
3
/h
4 m
3
/h
Actual
Velocity
0.071 m/s
0.142 m/s
0.283 m/s
0.566 m/s
Flowmeter
flowrate
0.530 m
3
/h
0.983 m
3
/h
2.046 m
3
/h
4.176 m
3
/h
Flowmeter
Velocity
0.075 m/s
0.139 m/s
0.289 m/s
0.591 m/s
1
2
3
4
Speed
0.071
0.142
0.283
0.566
Revise
0.943 1.017
0.978
0.958
ENT
Revise
NO
YES
Revise
Flow accuracy curve correct Factor,11 points
Choose
YES
and shift to
11points revision.
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
Содержание AMC2100 Series