3.9 System Mode
ENT
Sensor NO.
A
7
082106
Sensor No.
Press
button to move, Press
button to change the digit.
Size (mm)
0
065
Diameter
Press
button to move, Press
button to change the diameter
Sensor Factor
0
0.94100
Sensor Factor KS
KS is the original factory rectified factor, please don’t change it unless professional or technical
staff Press
button to move, Press
button to change the value
Output Check
NO
YES
Output Calibration
Press
or
button to choose
YES
select if needs calibration
Choose
NO
jump to Empty Tube Trim
Loop Test
4
8 12
16 20
Pulse Test
1 10 100
1000
2000
5000
4-20 mA simulated output
Press
or
button to select the simulated output value
4 . 8 . 12 . 16 . 20
Password
1111
Password login window
Input the system password:
Password: 1111
OPERATIONS
SET
ZERO ESC
ENT
Pulse Simulated output
Press
or
button to select the calibration type
1 . 10 . 100 . 1000 . 2000 . 5000
Coil Test
ON
X
Y Y
X
Excit voltage calibration
Check the coil X to Y or Y to X excit signal are correct or have fail
Press
or
button to select the calibration type
Choose X->Y, the Voltage will be 24VDC,
Choose Y->X, then Y to X will be 24 VDC
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
In the normal display, press
ENT
button for 5 seconds, then it will go to next setting
In this setting, Press
or
button till you selected SET
Revise
NO
YES
Revise
If choose YES, 11 points liner revision will be started, please see page 22-23 for details.
Choose
NO
jump to Output Calibration.