19
3.10 Advance Mode
Sensor NO.
A
7
082106
Size (mm)
0
065
Sensor Factor
0
0.94100
Output Check
NO
YES
Loop Test
4
8 12
16 20
Pulse Test
1 10 100
1000
2000
5000
Password
****
OPERATIONS
SET
ZERO ESC
ENT
Coil Test
ON
X
Y Y
X
Password login window
Input the system password:
Password: * * * *
Sensor No.
Press
button to move, Press
button to change the value
Diameter
Press
button to move, Press
button to change the diameter
Sensor Factor KS
KS is the original factory rectified factor, please don’t change it unless professional or technical
staff
Press
button to move, Press
button to change the value
4-20 mA simulated output
Press
or
button to select the calibration type
4 . 8 . 12 . 16 . 20
Pulse Simulated output
Press
or
button to select the calibration type
1 . 10 . 100 . 1000 . 2000 . 5000
Excit voltage calibration
Check the coil X to Y or Y to X excit signal are correct or have fail
Press
or
button to select the calibration type
Choose X->Y, the Voltage will be 24VDC, Choose Y->X, then Y to X will be 24 VDC
Output Calibration
Press
or
button to choose
YES
select if needs calibration
Choose
NO
jump to Empty Tube Trim
ENT
ENT
ENT
ENT
ENT
ENT
In the normal display, press
ENT
button for 5 seconds, then it will go to next setting
In this setting, Press
or
button till you selected SET
ENT
ENT
Revise
NO
YES
Revise
If choose YES, 11 points liner revision will be started, please see page 22-23 for details.
Choose
NO
jump to
Output Calibration
ENT