Meter Programming

IMPORTANT: The meter must be in Standby to 
begin any programming function. See Figure 1.

Change Unit of Measure

  1. Press and hold the 

Cal

 and 

Reset

 buttons at the same time.

• The 

Unit

 screen appears.

  2. Repeatedly press the 

Reset

 button.

• The screen scrolls through the available units of measure 

(see 

Figure 1

).

Once the desired units of measure appear:

  3. Press and hold the 

Cal

 button.

• The chosen units of measure are stored in memory and 

active.

• The meter cycles to Standby.

Calibration Factor Determination

To determine the meter’s calibration factor* in use:

  1. Press and hold the 

Cal

 button.

• The calibration factor screen appears.

The word 

Fact

 (short for factory) indicates the calibration set 

at the factory is in use (1.000). 

User

 indicates the 

Fact

ory

 

calibration factor has been superseded and is in use.

If the meter indicates 

User

 but the 

Fact

ory

 

calibration factor 

(1.000) is required:
  2. Press the 

Reset

 button.

• The screen changes from the 

User

 calibration factor to the 

Fact

ory

 

calibration factor.

  3. Press the 

Cal

 button.

• The meter cycles to Standby

.

• The 

Fact

ory calibration factor (1.000) is active.

• The 

User

 calibration factor is deleted from memory.

* The calibration factor is a value that the meter uses to 

calculate the amount of fluid dispensed.

User Calibration

Variables that cause a meter to require recalibration are:

• fluid viscosity
• fluid flow rate
• back-pressure

Field Calibration

This method of programming the meter requires dispensing a 

known quantity of fluid.

  1. Press and hold the 

Cal

 button.

• The calibration factor screen appears.

  2. Press and hold the 

Reset

 button.

• The 

Field

 screen appears.

  3. Dispense any amount of fluid greater than 5 quarts or liters 

into an appropriate-sized graduated beaker. 
• Dispense the fluid at the normal flow rate.

Should the field value match the dispensed value:

  4. Press the 

Reset

 button.

• An upward arrow appears.

  5. Press and hold the 

Reset

 button.

• The meter cycles to Standby.

The calibration factor remains the same.

If the field value does not match the actual amount:

  6. Press the 

Reset

 button.

• An upward arrow appears.

This arrow enables the user to increase the field value.

  7. Press the 

Reset

 button again to change the direction of the 

arrow.
• A downward arrow appears.

This arrow enables the user to decrease the value.

With the proper arrow chosen:

  8. Press the 

Cal

 button.

• The value changes one digit.

NOTE

: Press and hold the 

Cal

 button to change 

the value at a faster rate.

Once the required value displays:

  9. Press and hold the 

Reset

 button.

• The 

User

 calibration factor is stored in memory and active.

• The meter cycles to Standby.

Direct Calibration

This method of calibration is useful to correct an inaccuracy 

of a known percentage. 

For example, if the calibration factor is set at 1.000 and the 

meter registers more than the actual amount dispensed by 5%:

• the calibration factor should be decreased to 0.950

(1.000 x 0.95 = 0.950). 

Conversely, if the meter registers less that the actual amount 

dispensed by 5%:

• the calibration factor should be increased to 1.050

(1.000 x 1.05 = 1.050). 

To program a calculated calibration factor into the meter:

  1. Press and hold the 

Cal

 button.

• The calibration factor screen appears.

  2. Press and hold the 

Reset

 button.

• The 

Field

 screen appears.

  3. Press and hold the 

Reset

 button once again.

• The 

Direct

 screen appears with an upward arrow.

This arrow enables the user to increase the direct value.

  4. Repeat steps 

7

 through 

9

 above.

Содержание 3679

Страница 1: ...y volume Current delivery RESET TOTAL momentary TOTAL All three amounts display with a floating decimal point Current delivery 0 000 to 99999 RESET TOTAL and TOTAL 0 0 to 999999 The meter s non volatile memory retains the unit of measure totals and calibration factor during battery replacement Operation To zero the current delivery and RESET TOTAL the meter must be in Standby See Figure 1 Press th...

Страница 2: ...the Reset button The Field screen appears 3 Dispense any amount of fluid greater than 5 quarts or liters into an appropriate sized graduated beaker Dispense the fluid at the normal flow rate Should the field value match the dispensed value 4 Press the Reset button An upward arrow appears 5 Press and hold the Reset button The meter cycles to Standby The calibration factor remains the same If the fi...

Страница 3: ...nance procedure Repairs should only be performed by a qualified person using original repair parts Read each step of the instructions carefully Make sure a proper understanding is achieved before proceeding Battery Replacement NOTE All meter values remain in memory The batteries should be replaced once the battery icon appears on the display See Figure 1 IMPORTANT The meter will no longer register...

Страница 4: ...minals Meter does not measure accurately Incorrect calibration factor Recalibrate the meter Display blank Batteries installed incorrectly Install the batteries as indicated on the bottom of the meter Reduced or zero flow 1 Gears jammed 2 Clogged system 1 Overhaul metering gear cavity 2 Clean system filter Meter does not count and the flow rate is normal 1 Gears installed incorrectly 2 Register def...

Страница 5: ...se muestran con una coma decimal flotante Suministro actual de 0 000 a 99999 RESET TOTAL y TOTAL 0 0 a 999999 La memoria no volátil del medidor retiene la unidad de medida los totales y el factor de calibración durante el reemplazo de la batería Funcionamiento Para poner a cero el suministro actual y RESET TOTAL restablecimiento total el medidor debe estar en espera Consulte la Figura 1 Presione e...

Страница 6: ... Presione y mantenga presionado el botón Reset restablecimiento Aparece la pantalla Field campo 3 Surta cualquier cantidad de líquido mayor que 5 litros o cuartos de galón en un vaso de precipitación graduado de tamaño adecuado Surta el líquido según la tasa de flujo normal Si el valor del campo coincide con el valor que se surte 4 Presione el botón Reset Aparece una flecha hacia arriba 5 Presione...

Страница 7: ...izar cualquier procedimiento de mantenimiento Las reparaciones las deberán realizar sólo personal cal ificado utilizando repuestos originales Lea cuidadosamente cada paso de las instrucciones Asegúrese de comprenderlas antes de proseguir Reemplazo de la batería NOTA Todos los valores del medidor perman ecen en la memoria Las baterías se deben reemplazar una vez que el icono de batería aparezca en ...

Страница 8: ...actor de calibración es incorrecto Vuelva a calibrar el medidor La pantalla está en blanco Las baterías se instalaron en forma incorrecta Instale las baterías como se indica en la parte inferior del medidor El flujo es reducido o inexistente 1 Los engranajes están atascados 2 El sistema está obstruido 1 Revise la cavidad del engranaje de medición 2 Limpie el filtro del sistema El medidor no funcio...

Страница 9: ...ané TOTAL Les trois totaux sont affichés avec un signe décimal flottant Distribution actuelle 0 000 à 99999 RESET TOTAL et TOTAL 0 0 à 999999 La mémoire non volatile conserve l unité de mesure les totaux et le facteur de calibration lors du remplacement des piles Mode d emploi Pour remettre à zéro la distribution actuelle et RESET TOTAL total réglable le compteur doit être en mode d attente Se rep...

Страница 10: ...an Field zone est affiché 3 Distribuez une quantité de fluide de plus de 5 litres ou pintes dans un bécher gradué de taille appropriée Distribuez le fluide à un débit normal Si la valeur de la zone correspond à la valeur distribuée 4 Appuyez sur le bouton Reset Une flèche pointant vers le haut est affichée 5 Appuyez et maintenez enfoncé le bouton Reset Le compteur manoeuvre en position d attente L...

Страница 11: ...ys tème avant d effectuer toute procédure d entretien Toutes réparations doivent être réalisées par un person nel qualifié et en utilisant des pièces de rechange d orig ine Lire chaque étape des instructions avec soin S assurer de bien comprendre toute la procédure avant de com mencer Remplacement des piles REMARQUE toutes les valeurs du compteur sont gardées en mémoire Les piles doivent être remp...

Страница 12: ...on Facteur de calibration incorrect Recalibrer le compteur Affichage vide Piles installées incorrectement Installer les piles tel qu indiqué dans la partie inférieure du compteur Débit réduit ou nul 1 Engrenages bloqués 2 Système encrassé 1 Remettre en état la cavité d engrenage de compteur 2 Nettoyer le filtre du système Le compteur ne compte pas et le débit est normal 1 Engrenages installés inco...