NOVA 120 Pointing model user manual
42
All the figures of IH, ID… DCES are not the same, but are close, or have opposite signs. MA and ME are directly related to
polar alignment errors and must be the same. It must be noticed that polar alignment of the mount used to create the
document is really excellent!
MA is jumping from -10.4 arcsec to 1.4 arcmin error, ME from -2.5 arcsec to -1.2 arcsec. This denotes a slight pillar
flexure when tube is located at the West side of the pier and the East side. This is why for German mount, performing
two models is always employed by professionals.
Pointing model is really powerful to extract what is happening to the whole setup (tel mount) and sometime
can help to apply fixes: polar alignment being too bad, tube flexion and so on.
3. Applying the pointing model to the NOVA mount
Once the two
.cor
files are generated by the process above, they need to be entered into the NOVA mount software.
First, close PRISM’s connection to the NOVA mount software.
Then to the NOVA mount software, go to setup then open the pointing model setup form:
Enable pointing model and select the
TubeWestPier-10-10-14.cor
and
TubeEastPier-10-10-14.cor
accordingly. So having
named the files this way is a good trick to avoid mixing up these files.