Depending on the sensor calibration drift, up to 4 samples may be required.
• If only 2 tests are required, a Verification OK message is displayed when thecalibration
is complete.
• If 3 or 4 tests are required, a Calibration OK message is displayed when calibration is
complete.
13.
Press the right button to Finish (there is also the option to retry by pressing the left button).
Calibration is complete.
The V3 handset returns to ready state and Blow for 5 seconds is displayed.
NOTe! If the calibration fails on numerous attempts, the sensor may need
replacement. Contact ACS for return instructions.
TROUbLESHOOTING
In the event of a calibration failure, first check these common causes of errors:
• Plastic tubing has been overused, or has condensation present.
• The alcohol gas standard value is not 130PPM, or the value has been incorrectly entered in the
Select Value menu.
• The alcohol gas standard is expired.
• A connection to or from the gas valve is not secured properly (check all connections).
• The value entered in Altitude menu is incorrect.
ALTITUdE REFERENCE GUIdE
The altitude adjustment is in increments of 200 meters. Choose the closest city elevation and
round the altitude up or down to the closest value accordingly.
ExApMLE
CITY
ALTITUdE (M)
SETTING
Amsterdam
0
0
Berlin
34
0
Brussels
48
0
Geneva
373
400
Madrid
654
600
Moscow
189
200
Ottawa
79
0
Paris
34
0
Rome
14
0
Washington
7
0