CYLINdER SETUp

1. 

Examine the canister and valve for any damage, and check the expiration date on the label. 

Observe all cautions and safety warnings found on the canister.

2. 

Remove the protective cap and screw the regulator valve to the canister. Attach the rigid 

delivery tube. Never modify the delivery tube in any way.

CAUTION! Damaged or broken valves can turn a canister into an unguided missile. 

Attach the valve in a safe location.

CALIbRATING A HANdSET

1. 

Power the handset on by holding down either the right or left button.

2. 

Access the Menu by pressing and holding the left button. Scroll through the menu and select 

System Maintenance. The four digit code which is provided with the V3 accesses this 

menu.

3. 

Enter the code and move the cursor to OK? when finished. Push the right button to Select. 

This enters the Service menu.

4. 

Scroll through the Service menu and press select Calibration.

5. 

The current handset Date/Time is displayed. To proceed select OK with the right button. This 

enters the Select Type menu.
NOTe! If a tachograph is installed the date/time cannot be changed.

6. 

The Select type menu contains the calibration options. Scroll to dry Gas and push the right 

button to Select. This enters the Select value menu.

7. 

The Select value menu contains the gas standard concentration options. Verify that 130PPM 

is on the canister label, and use the left button to scroll to 130ppM. Push the right button to 

Select. This enters the Altitude menu.

8. 

The Altitude menu contains the altitude correction factor. Verify the altitude value using the 

altitude reference guide, which is on the last page of this guide. The left button increases in 

increments of 200 meters. Press the right button to Accept.

9. 

Wait message is displayed followed by a 2:00 minute warm-up. The handset is ready for 

calibration when the Turn on gas message is displayed.

10. 

Insert the delivery tube from the regulator valve into the mouthpiece port. Keep hold of the 

handset as the delivery tube is rigid.

11. 

Press the valve down to release the gas while keeping 

hold of the handset (see figure). The handset analyzes 

the gas standard and displays a blowing… message 

followed by Analyzing message.  

 

Continue holding the valve down until the handset clicks 

and then beeps, indicating that the test is complete.

12. 

The handset analyzes the test, which is 

followed by a Wait message and 0:45 second 

countdown. The handset is ready for the second sample 

when the Turn on Gas message is displayed. 

 

 

Содержание V3 B-2 Series

Страница 1: ...Service Guide V3 SERIES B 2 ...

Страница 2: ...n this document is the valuable property of Alcohol Countermeasure Systems Corp and all copyright and other proprietary rights to this document are reserved No reproduction of this document is permitted without the prior written consent of Alcohol Countermeasure Systems Corp ALCOLOCK ALCOHOL COUNTERMEASURE SYSTEMS and the Molly are trademarks of Alcohol Countermeasure Systems International Inc and...

Страница 3: ... system To test that all inputs to the V3 are off and that both pushbuttons are working if the handset is awake hold both pushbuttons for 3 seconds The handset should go to the lower power state the display and all the LEDs should turn off To test the display backlights and left pushbutton Wake the handset up by pressing the left pushbutton The display should show Alcolock Disabled OK to start eng...

Страница 4: ...eceived from the tachograph over the CAN BUS 1 Check if tachograph is functioning properly 2 Check wiring from tachograph to ECU 3 Replace ECU Communication error Handset has lost communication with ECU 1 Check voltage of vehicle during ignition cranking Voltage should stay above 9V or 16V depending on ECU 2 Replace spiral cable 3 Replace ECU 4 Replace Handset ECU voltage incorrect ECU is connecte...

Страница 5: ...ace ECU Calibration Procedures It is recommended calibration be done indoors in a service facility using an ACS download station The calibration procedures are intended for use by trained technicians only Equipment Needed 130PPM 0 05 BAC Alcohol gas standard ACS 95 000425 Regulator valve assembly 6L min flow rate ACS 94 000225 V3 Download station ACS 79 006111 Compressed Gas Safety Calibration sho...

Страница 6: ...n options Scroll to Dry Gas and push the right button to Select This enters the Select value menu 7 The Select value menu contains the gas standard concentration options Verify that 130PPM is on the canister label and use the left button to scroll to 130PPM Push the right button to Select This enters the Altitude menu 8 The Altitude menu contains the altitude correction factor Verify the altitude ...

Страница 7: ...s Troubleshooting In the event of a calibration failure first check these common causes of errors Plastic tubing has been overused or has condensation present The alcohol gas standard value is not 130PPM or the value has been incorrectly entered in the Select Value menu The alcohol gas standard is expired A connection to or from the gas valve is not secured properly check all connections The value...