Alcolizer WM4 Скачать руководство пользователя страница 12

 

 
 

 
REF: WM4 User Manual V7.docx   

 

 

Page 12    

 

 

© Alcolizer Technology

 

 

3.2

 

Mounting and Starting the instrument 

3.2.1

 

Mounting 

Mount the instrument in accordance with the following procedures: 

a.

 

Using an assistant, hang the instrument onto the back plate by aligning the 
male lift off hinge pins on the back plate with the main instrument lift of 
female hinge tubes. 

b.

 

Confirm the instrument swings effectively and does not impact any other 
fixtures. 

c.

 

Route any cables to avoid damage and trip hazards. 

d.

 

Visually check that there are no loose components from transportation. 

e.

 

Remove any packing materials. 

f.

 

Remove the screen protector film from the front of the display screen.  

g.

 

Load straws into the straw box, ensuring that the straws are lying freely 
within the box. 
 
 

 

3.2.2

 

Initial starting 

Perform the following steps for the initial start-up: 

a.

 

Turn the reference gas on by rotating the on/off valve one full turn anti-
clockwise  

 

WARNING 

Do not use adhesives (Liquid Nails) as the method of fixing as these are not considered to be a 

reliable or safe fixing means for the WM4. 

NOTE 

The gas bottle certification documents (attached to the gas bottle neck)  

must remain with the bottle. 

Содержание WM4

Страница 1: ...btained by correct use of this device is only considered accurate at the time of testing Great care has been taken to ensure the accuracy of each reading Neither the manufacturer the distributor nor the owner accepts liability or responsibility due to any action or claim arising from the reading produced by this device whether used correctly or incorrectly ...

Страница 2: ...Mounting and Starting the instrument 12 3 2 1 Mounting 12 3 2 2 Initial starting 12 4 Operating Instruction 13 4 1 Providing a breath sample 13 4 2 Analysing the breath sample 15 4 3 Invalid sample 15 5 Owner Servicing 16 5 1 Straw Replenishment 16 5 2 Cleaning 17 6 User Controls 18 6 1 Buttons 18 6 2 Keyboard 19 7 User Interface 20 7 1 Accessing the user menu 20 7 2 Status 21 7 2 1 Identification...

Страница 3: ...uration 29 11 1 1 Network Setup 29 11 1 2 GSM Set up 30 11 2 Services 30 11 2 1 Authorisation Over Network 30 11 2 2 Alcolizer Portal 30 12 Email 31 12 1 Email Settings 31 12 1 1 Settings Mail template 31 12 1 2 Settings Mail server 32 12 1 3 Test Mail 32 12 2 Email Reports 33 12 2 1 Breath Test Results 33 12 2 2 Instrument Status 33 12 2 3 Event Log Report 33 13 Relays 34 13 1 Hardware 34 13 2 Co...

Страница 4: ...tocol 38 16 2 Wiegand USB device 38 16 2 1 Direct Network Connection 38 16 2 2 USB Input Device Authorisation 38 16 3 Coin Mechanism Authorisation 38 17 Factory Servicing and Troubleshooting 39 17 1 Factory Servicing 39 17 2 Field Servicing 39 17 3 Troubleshooting 39 18 Calibration 41 ...

Страница 5: ...le All variants have internet connectivity through an Ethernet port or GSM module in the case of the cellular version Some variants include a mix of the types listed above Depending on the specific user requirements machines can include unique configurations Should any concerns over the configuration be noted contact your Alcolizer Technology representative for further advice WARNING If the supply...

Страница 6: ...kg Power 100 240V 50 60Hz 75W Housing Steel construction Sensor Alcohol specific electrochemical fuel cell Response Time Within 3 5 seconds for a 0 100g 100ml BAC reading Display Type 25cm 10 colour graphics LCD Operating Temp 0 to 50 c Measurement Display 4 Digits x xxx g 100ml Detection Range 0 000 0 500 g 100ml Accuracy Absolute accuracy to better than 0 005 g 100ml BAC Sample volume 1 2 Litres...

Страница 7: ...int commercial and industrial alcohol breath testing applications The Alcolizer WM4 incorporates innovative and pioneering technology with features providing speed and efficiency for all testing requirements It incorporates a self recalibration feature providing accuracy and reliability from its own internal NATA Certified Gas Standard every 24 hours 10 Display screen Straw receptacle sample port ...

Страница 8: ...REF WM4 User Manual V7 docx Page 8 Alcolizer Technology Serial number decal Serving data decal Key lock Information decal ...

Страница 9: ...REF WM4 User Manual V7 docx Page 9 Alcolizer Technology ...

Страница 10: ...main unit Avoid mounting the instrument above anything that could be affected by moisture e g power points The instrument must not be exposed to extreme temperatures outside of the defined operating range 3 1 2 Mounting height and position The instrument must be mounted and positioned at a height to accommodate the majority of users and in such a way to allow access to the key lock and to allow th...

Страница 11: ...the back plate from the unit b Choose a suitable mounting site as described previously c Bend or remove any cutaway portions on the back plate to permit cable access if necessary d Fix the back plate to the wall using masonry fixing hardware e g medium duty anchors The fixings should be strong enough to support the instrument at least 20kg and any incidental loading such as a person leaning or pul...

Страница 12: ...zards d Visually check that there are no loose components from transportation e Remove any packing materials f Remove the screen protector film from the front of the display screen g Load straws into the straw box ensuring that the straws are lying freely within the box 3 2 2 Initial starting Perform the following steps for the initial start up a Turn the reference gas on by rotating the on off va...

Страница 13: ...and instructions on how to provide a breath sample There may be a specific customer welcome screen b Take a straw from the straw holder c Place straw in the sample port NOTE Food or drink must not be consumed for the 10 minutes prior to testing The BAC can continue to rise for up to two hours after the cessation of drinking care should be taken if a result close to the designated BAC limit is indi...

Страница 14: ...colizer Technology d Start blowing and continue blowing until the progress bar reaches 100 During the blow sequence a beeping sound will be heard e Stop blowing when the display shows Stop Blowing This will also be indicated by a long beep ...

Страница 15: ... 050g 100ml the display will be red The result will continue to be displayed while the instrument resets itself ready for the next test For zero readings this will typically be about five seconds 4 3 Invalid sample The invalid blow screen is displayed if the user stops blowing before 1 2 litres is recorded they blow too softly blow too hard or for not long enough If the blow is invalid cease blowi...

Страница 16: ...nging it open c Open the straw bin by lifting the latch and removing the cover by manoeuvring it forward d Load in new straws ensuring they lie loosely and tidily in the bin e Reinstall the bin lid by aligning it in the rear slots and latching it down Ensure it is closed securely f Close and lock the instrument g Turn the power back on The instrument will conduct internal checks and then be ready ...

Страница 17: ...aterials abrasive cleaning pads or chemicals Non alcohol based cleaning wipes can be used to clean the Wall Mount 4 As part of routine cleaning use the brush provided to clean out the Wall Mount sample port 1 Insert the bush bristles into the sample port 2 Do Not exceed the length of the bristles through the sample port 3 Continue a backward forward motion with the brush for about 5 minutes 4 Test...

Страница 18: ...hat users need to operate are FN1 FN4 and OFF Do not touch the other buttons The buttons have the following functions FN1 Show hide the Status page Does the same as pressing F1 on the keyboard FN4 Save the event log to USB flash memory Does the same as pressing F4 on the keyboard OFF Power off and power on the instrument If the instrument is running press once to power down If the instrument is po...

Страница 19: ...ts located on the main circuit panel Once the keyboard is connected access to the menu is available The keyboard functions are as follows WARNING Pressing and holding the OFF button may cause irreparable damage to the instrument and should only be performed by an Alcolizer Technician WARNING When the unit is open or not installed on the back plate electrical components are exposed The following st...

Страница 20: ...Interface 7 1 Accessing the user menu To access the User Menu perform the following steps a Connect a keyboard in accordance with paragraph 0 b Press the F1 key to access the WM4 User Menu area c Press the left Alt key to bring up available menu choices d Press the down key and move down to the required menu e g Settings e Press left Alt to hide the menu and give clear access to the selected page ...

Страница 21: ...nalysis is disabled if an alert is active Warning The instrument has discovered a problem but is able to function normally for the time being For example calibration gas low Breath analysis is enabled if a warning is active Notice Information that might be useful to a technician but in no way represents a problem with the instrument Breath analysis is enabled if a notice is active Disabled Breath ...

Страница 22: ...STAT5 displays the WM4 State The LEDs various displays indicate the following STAT4 LED Off Log save not in progress LED On Log save triggered or in progress STAT5 LED fast blinking Reset LED two short pulses Starting up LED continuous one short pulse Running LED steady blinking Shutting down LED on Shut down power still connected and On LED off Unit Off If any other blinking patterns are noticed ...

Страница 23: ...ect date and time the time zone must be set correctly The display format is set to YYYY MM DD HH MM SS and cannot be changed 8 4 Authorisation The authorisation feature when enabled prevents the WM4 from performing a breath test until it has received authorisation from an external agent e g remote serial device keypad Coin Mechanism etc Enabled Check to enable breath test authorisation For further...

Страница 24: ...ow password characters entered on the screen should be displayed 8 6 Shutdown Instrument When the Shutdown Instrument button is pressed the following options are presented to the user Reboot Restart the instrument Power Off Shutdown the instrument so that mains power can be removed This method or the OFF button on the PCBA localised internally in the WM4 are the only two ways to correctly power of...

Страница 25: ...User Menu in accordance with the procedures at paragraph 7 1 b From the User Menu access the custom appearance screen by scrolling down to Custom Appearance and pressing space bar The Custom Appearance screen will appear c Tick the Show Idle Background Image box d Inset a USB storage device thumb drive with the desired image into an available USB port e Click on the Select Image button f Select th...

Страница 26: ...c Tick the Use Custom Text box in the elements field d Select the desired screen e Type the required text in the free text field for the highlighted screen f To activate a free text of a specific screen tick the box beside the screen name 9 4 Intervals It is possible to adjust the timings for some screen display intervals Descriptions of how the timings apply to the display are presented for each ...

Страница 27: ... shut down to be powered off or rebooted System Started The instrument was started System Time Changed The wallmount s time or date was changed Fault Cleared A fault status item cleared Fault Set A fault status item set Alert Cleared An alert status item cleared Alert Set An alert status item set Warning Cleared A warning status item cleared Warning Set A warning status item set Notice Cleared A n...

Страница 28: ...g a USB flash drive and pressing FN4 on the PCBA Refer to paragraph 6 1 to see button location These are the formats which logs can be exported CSV Comma Separated Values with one log record per row and record fields separated by commas within rows This is suitable for import into most spreadsheet applications JSON JavaScript Object Notation in a format suitable for integration with automated syst...

Страница 29: ...net connection as follows a Connect the WM4 to an internet receptacle via the Ethernet cable b Connect a keyboard mouse and access the User Menu in accordance with the procedures at paragraph 7 1 c From the User Menu access the network configuration screen by scrolling down to Network and pressing space bar The Network Configuration page will appear d Complete the Configuration page with the requi...

Страница 30: ...a TCP IP port to allow breath test authorisation over the network The operation of this feature is identical to the protocol operating over the serial ports There is no limit to the number of network clients that can connect simultaneously Enabled Whether the instrument accepts network connections on the specified port This is disabled by default Port The TCP port number 11 2 2 Alcolizer Portal Th...

Страница 31: ... screen has two tabs Settings and Reports 12 1 1 Settings Mail template The Mail Template panel allows configuration of the format and recipients of emails sent from the instrument Default Recipients A list of email recipients for emails produced by the instrument e g someone bigmine com au Add additional recipients as required separating each with a space Subject Prefix Text to be prepended to th...

Страница 32: ...orarily display the password characters Hostname URL of the SMTP server e g smtp gmail com Port The TCP IP port number used by the SMTP server Leave blank to use a default value Use SSL If enabled then use secure sockets layer to connect to the SMTP server 12 1 3 Test Mail The Test Email panel allows sending a test email from the system confirm email settings and provide feedback on any issues the...

Страница 33: ... information on all currently active status items not just the status item that changed state The status options are Faults Permanently Out of Order send an email if a fault is raised or cleared Alerts Temporarily Out of Order send an email if an alert is raised or cleared Warning send an email if an alert is asserted or cleared Notice send an email if a notice is asserted or cleared 12 2 3 Event ...

Страница 34: ... each other The minimum and maximum values can overlap such that one two or three relays can be triggered simultaneously If a new breath test result is calculated while a relay output is asserted from the previous test the relay takes the value required by the new test 13 1 Hardware The relays outputs are labelled on the circuit board as Relay1 Relay2 and Relay3 Each relay output has three termina...

Страница 35: ...m value is 0 00 the relay should turn on if the value of the result is less than or equal 0 00 Duration If the relay is enabled and a breath test result falls within its configured range the relay should fire for a number of seconds equal to the configured duration The Relay Status indicates if the relay is currently ON or OFF There are three buttons on this page that briefly activates the relays ...

Страница 36: ...validate the instrument configuration and behaviour Each test indicates whether it has been recently tested and lists a title as well as provides a description of what the test does One many or all tests can be run Some tests require a degree of user interaction and if this occurs instructions will appear on the screen indicating how to proceed ...

Страница 37: ...logged with the breath test result in the event log and the wallmount waits for authorisation to perform further tests If a breath test is not performed within a timeout period the authorisation expires and the wallmount must receive a new authorisation code to perform further tests Authorising agents connect to the wallmount through authorisation channels An authorisation channel is a combination...

Страница 38: ...onfigurable port number When a remote device connects to the wall mount it will receive all breath test results according to the americium protocol specification If breath test authorisation is enabled the remote device can also authorise breath tests There is no limit to the number of devices that can connect simultaneously to the wall mount 16 2 2 USB Input Device Authorisation Any HID input dev...

Страница 39: ...by turning the valve on the gas cylinder d Remove straws and tape down straw bin lid e Using two man lifting techniques lift the instrument off the back plate hinges to and remove it from the wall f Pack the instrument into suitable packaging to ensure it is not damaged in freight 17 2 Field Servicing Contact Alcolizer Technology for further information 17 3 Troubleshooting Should the instrument m...

Страница 40: ...24 hours The instrument continues to display an Out of Order message Contact Alcolizer Technology for further advice 1300 789 908 The instrument is displaying a warning icon in the bottom left corner Either there are status items requiring attention press FN1 to identify items requiring attention or the service life may shortly be about to expire Contact Alcolizer immediately to arrange annual ser...

Страница 41: ...quid is not present and try again If the problem persists allow a suitable time period for the liquid to clear itself anywhere up to 48hrs or refer to section 5 2 of this manual If the problem persists contact Alcolizer on 1300 789 908 18 Calibration The WM4 self calibrates every 24 hours for the duration of its service life and requires no user calibration Liquid Detected ...