Centurion Quantum
User Manual
REF: Centurion Quantum User Manual V2.docx
Page 20
© Alcolizer Technology
Uncontrolled when printed.
7.2.1
Instrument
Serial Number:
A number that uniquely identifies the Centurion Quantum. This
number is also printed on the side panel safety label located on the side of the
instrument.
Software Version:
The version of software running in the Centurion Quantum (e.g.,
fm-17.0).
7.2.2
Alcohol Sample Module
Serial Number:
A number that uniquely identifies the Alcohol Sample Module.
Application Version:
The version of software running in the Alcohol Sample Module
(e.g., 2.06).
Days to Calibration:
Total days until the next calibration is due.
7.2.3
Status Items
Each status item is classified as one of the following.
Fault:
The instrument has encountered a critical error and is unable to perform
breath analysis. The fault must be manually cleared (or repaired) by an authorised
Alcolizer service technician on
1300 789 908.
Alert:
The instrument has encountered a temporary condition that is preventing the
unit from performing breath analysis. The alert will clear automatically. Breath
analysis is disabled if an alert is active.
Warning:
The instrument has discovered a problem but is able to function normally
(for the time being).
Notice:
Information that might be useful to a technician but in no way represents a
problem with the instrument. Breath analysis is enabled if a notice is active.
Disabled:
Breath testing is disabled for a reason other than a fault or alert.
7.2.4
Network Connections
The Network Connections area shows status information for the Centurion Quantum
Ethernet port.
A valid network connection is not required for the instrument to operate correctly,
however if a Centurion has an internet connection, the date and time will be
automatically adjusted.
If connected to a valid network the instrument's IPv4 address is displayed (e.g.,
192.168.15.133).
If not connected to a valid network,
none available
is displayed.
NOTE
Status items are displayed to the user on the Status page in the instrument's current
language. Each status item also includes a unique code to help identify the status item
when the instrument is in a different language.
WARNING
When the instrument is open or not installed on the back plate, electrical components are
exposed. The following steps require the user to open the Centurion with power applied.
Do not come in contact with any exposed components.