ATS9462 User Manual
53
Bandwidth 65
MHz
Trigger Delay
Software selectable from 0 to 9,999,999
sampling clock cycles
Trigger Timeout
Software selectable with a 10 us resolution.
Maximum settable value is 3,600 seconds.
Can also be disabled to wait indefinitely for a
trigger event
EXT (External Trigger) Input
Input impedance
1 M
Ω
in parallel with 30pF
±
10pF
Bandwidth (-3dB)
DC-coupled
DC - 25 MHz
AC-coupled
10 Hz - 25 MHz
Input range
±
5V or
±
1V, software selectable
DC accuracy
±
10% of full-scale input
Input protection
±
28V (DC + peak AC without external
attenuation)
Coupling
AC or DC, software selectable
Auxiliary I/O (AUX I/O)
Signal Direction
Input or Output, software selectable
Output by default
Output Types:
Trigger Output
Busy
Output
Software controlled Digital Output
Input Types:
Trigger Enable
Software readable Digital Input
Output
Amplitude:
5 Volt TTL
Synchronization:
Synchronized
to rising edge of sampling clock
Input
Amplitude:
5 Volt TTL or 3.3 Volt TTL
Certification and Compliances
CE Mark Compliance
Materials Supplied
One ATS9462 Digitizer
One ATS9462 Install Disk
One ATS9462 User Manual
All specifications are subject to change without notice