INTRODUCTION

Thank you for purchasing DOH-SD1 Dissolved Oxygen Meter with Datalogging SD Card.
Please read this user’s manual carefully and thoroughly before using the instrument.

Many commercial and industrial processes carefully monitor the concentration of oxygen
contained in water because dissolved oxygen (DO) level is an excellent indicator of water
quality. For example, if the DO level of a body of water remains too high or too low for too long,
fish and other aquatic organisms will die or become more susceptible to disease. For quality
control purposes, many industries other than fish farming and water pollution control routinely
measure and keep track of the DO levels of their process streams in real time. Among them are
food processing, photography, pulp and paper manufacturing, power generation, electroplating
and facilities management.

The DOH-SD1 has all of the features needed by an instrument suitable for precision
measurement of DO levels in a wide range of applications. Its 7.5-inch-long, polarographic,
electrolyte-filled oxygen probe measures not only the DO concentration of a liquid, but the
liquid’s water temperature as well. The meter uses a single custom microcontroller to display
readings from 0 to 20 milligrams/liter (mg/L) with a resolution of 0.1 mg/L and an accuracy of ±
0.4 mg/L.

Because it is microprocessor-based, the DOH-SD1 can make full use of the portability, reliability
and large storage capacities that SD memory cards offer. Measurements can be made
automatically at any sampling rate between one second and nine hours. After the instrument
time-stamps and stores the measurements on an SD card plugged into it (a process called
datalogging), the user can remove the card and plug it into a laptop or desktop computer either
directly or via a USB card reader. The data logs are stored on the card as files with the .xls
extension that can be opened by Microsoft’s Excel application.

The DOH-SD1 has a backlit 1.5 x 2.0-inch display and is powered by six AA Alkaline batteries
or an optional 9-VDC AC adapter.

KEY FEATURES

 

Uses heavy-duty polargraphic oxygen probe

 

Precisely measures DO from 0 to 20 mg/L, as well as liquid temperature

 

Easy to calibrate

 

Supports automatic and manual datalogging

 

Time-stamps and stores measurements on SD memory cards

 

Sampling time settable from 1 second to 9 hours

 

Automatically compensates for altitude and salt content of liquids

 

1.5 x 2.0-inch backlit liquid crystal display

 

Powered by six AA batteries or optional 9-VDC adapter

 

Readings switchable between degrees F and C

2. SPECIFICATIONS

Содержание OMEGA DOH-SD1

Страница 1: ...DOH SD1 Disolved Oxygen Meter With Data Logging SD Card...

Страница 2: ...atic vs manual datalogging 9 Transferring data from the SD card to a computer 10 Specifications 12 Maintenance troubleshooting tips 13 Replacement parts accessories 14 1 FEATURES Office Jl Radin Inten...

Страница 3: ...microcontroller to display readings from 0 to 20 milligrams liter mg L with a resolution of 0 1 mg L and an accuracy of 0 4 mg L Because it is microprocessor based the DOH SD1 can make full use of th...

Страница 4: ...l the probe Two spare probe heads with diaphragm set A 2 GB SD memory card This user s manual Fig 1 shows all of the controls and indicators on the front back top bottom and side of the DOH SD1 Famili...

Страница 5: ...e Replace the cover by replacing the two screws 2 Install an SD card To prepare for setup also install the supplied 2 GB SD memory card or another card with a capacity from 1 GB to 16 GB in the socket...

Страница 6: ...current month and press the REC ENTER button to store the setting When the next screen flashes the value above dd again use the or button to navigate to the current day and press the REC ENTER button...

Страница 7: ...pper portion of the display matches your altitude in meters Then press the REC ENTER button to store this value as the temporary default 12 Compensate DO measurements for altitude in feet Once DO meas...

Страница 8: ...obe head with electrolyte To calibrate the DOH SD1 use the following procedure while referring to Figures 1 and 3 CALLOUTS FOR FIG 3 3 1 Probe Plug 3 2 Probe Handle 3 3 Temperature sensing Metal 3 4 P...

Страница 9: ...el once Insert the probe at least 4 inches 10 cm into the liquid to be measured Wait a few minutes for the instrument to automatically compensate for any difference in temperature between the probe an...

Страница 10: ...which the instrument was set up press the LOGGER SET button which has the words Sampling check stenciled below it once To change the sampling time perform Step 5 of the setup procedure 1 To start aut...

Страница 11: ...ation press the REC ENTER button to save the setting This causes the value to stop flashing and the P to return 6 To end manual datalogging press the REC ENTER button and hold it for at least three se...

Страница 12: ...11 5 Typical Excel graphics only screen 6 Typical Excel mixed graphics data screen...

Страница 13: ...Automatic Compensation for altitude 0 to 29 200 ft 0 to 8 900 meters Compensation for liquid salt content 0 to 50 Datalogging sampling time range 1 second to 8 hours 59 minutes 59 seconds DO probe typ...

Страница 14: ...t corner of the display it s time to replace the six AA batteries that power it although measurements will remain valid for several hours after the low battery indicator first appears Replacing the ba...

Страница 15: ...ACEMENT PARTS OPTIONAL ACCESSORIES DOE SD1 Spare oxygen probe DOEH SD1 Spare probe head and diaphragm set DOEL SD1 Electrolyte for probe APAPTER ST 9 VDC adapter for 110VAC power supply RS232 SD RS232...