ALAT UJI OMEGA DOH-SD1 Скачать руководство пользователя страница 10

9

Whenever the instrument is in measurement mode, you also can check the current date and time
by briefly pressing the TIME button. Doing so causes both values to appear briefly at the lower
left of the display.

Holding and storing measurements

1.

 

To hold a measured value

, press the HOLD/ESC button during the measurement. Doing

so will cause the word “HOLD” to appear at the top of the display. Pressing the
HOLD/ESC button again releases the hold.

2.

 

To record and recall readings

, press the REC/ENTER button while making

measurements. This will make the term “REC” appear at the top of the display. Pressing
the REC/ENTER button again, briefly, will make the term “MAX” appear to the right of
“REC” and switch the display to the maximum value stored in memory during the last
recording session. Pressing the REC/ENTER button again, briefly, will make the term
“MIN” appear to the right of “REC” and switch the display to the minimum value stored
during the last session.

3.

 

To exit recording mode

, press and hold the REC/ENTER button for at least three

seconds, until the term “REC” disappears from the top line of the display. The display
will then revert to showing the current reading.

Automatic vs. manual datalogging

The DOH-SD1 can automatically log data at a user-selected sampling period from 1 second to 9
hours. To view the sampling time for which the instrument was set up, press the LOGGER SET
button (which has the words “Sampling check” stenciled below it) once. To change the sampling
time, perform Step 5 of the setup procedure.

1.

 

To start automatic datalogging

, press the REC/ENTER button once. The top line of the

display will then show the term “REC”. Pressing the LOGGER SET button at this point
will make REC flash and add the flashing term “LOGGER” at the top right of the
display. This indicates that the instrument is currently storing measured values and their
time stamps in memory.

2.

 

To pause automatic datalogging

, press the LOGGER SET button once; this action

makes the flashing term LOGGER disappear from the top right of the display and
changes the term “REC” from flashing to constant. Pressing the LOGGER SET button
again resumes automatic datalogging.

3.

 

To end automatic datalogging

, press the REC/ENTER button and hold it for at least

two seconds. This action causes the “REC” message to disappear.

4.

 

To log data manually

, set the sampling time to zero using Step 5 of the setup procedure.

Then press the REC/ENTER button once. The display will show the term “REC” on the
top line, a value in the middle, and below it the letter “P” on the left and a number from 1
to 99 on the same line to the right. The number indicates the position on the SD card that
will be used to store manually logged data.

Now press the LOGGER SET button. This will cause the beeper to sound 

[even if disabled?

Engineering, pls check]

and the term “LOGGER” to briefly appear at the upper right of the

Содержание OMEGA DOH-SD1

Страница 1: ...DOH SD1 Disolved Oxygen Meter With Data Logging SD Card...

Страница 2: ...atic vs manual datalogging 9 Transferring data from the SD card to a computer 10 Specifications 12 Maintenance troubleshooting tips 13 Replacement parts accessories 14 1 FEATURES Office Jl Radin Inten...

Страница 3: ...microcontroller to display readings from 0 to 20 milligrams liter mg L with a resolution of 0 1 mg L and an accuracy of 0 4 mg L Because it is microprocessor based the DOH SD1 can make full use of th...

Страница 4: ...l the probe Two spare probe heads with diaphragm set A 2 GB SD memory card This user s manual Fig 1 shows all of the controls and indicators on the front back top bottom and side of the DOH SD1 Famili...

Страница 5: ...e Replace the cover by replacing the two screws 2 Install an SD card To prepare for setup also install the supplied 2 GB SD memory card or another card with a capacity from 1 GB to 16 GB in the socket...

Страница 6: ...current month and press the REC ENTER button to store the setting When the next screen flashes the value above dd again use the or button to navigate to the current day and press the REC ENTER button...

Страница 7: ...pper portion of the display matches your altitude in meters Then press the REC ENTER button to store this value as the temporary default 12 Compensate DO measurements for altitude in feet Once DO meas...

Страница 8: ...obe head with electrolyte To calibrate the DOH SD1 use the following procedure while referring to Figures 1 and 3 CALLOUTS FOR FIG 3 3 1 Probe Plug 3 2 Probe Handle 3 3 Temperature sensing Metal 3 4 P...

Страница 9: ...el once Insert the probe at least 4 inches 10 cm into the liquid to be measured Wait a few minutes for the instrument to automatically compensate for any difference in temperature between the probe an...

Страница 10: ...which the instrument was set up press the LOGGER SET button which has the words Sampling check stenciled below it once To change the sampling time perform Step 5 of the setup procedure 1 To start aut...

Страница 11: ...ation press the REC ENTER button to save the setting This causes the value to stop flashing and the P to return 6 To end manual datalogging press the REC ENTER button and hold it for at least three se...

Страница 12: ...11 5 Typical Excel graphics only screen 6 Typical Excel mixed graphics data screen...

Страница 13: ...Automatic Compensation for altitude 0 to 29 200 ft 0 to 8 900 meters Compensation for liquid salt content 0 to 50 Datalogging sampling time range 1 second to 8 hours 59 minutes 59 seconds DO probe typ...

Страница 14: ...t corner of the display it s time to replace the six AA batteries that power it although measurements will remain valid for several hours after the low battery indicator first appears Replacing the ba...

Страница 15: ...ACEMENT PARTS OPTIONAL ACCESSORIES DOE SD1 Spare oxygen probe DOEH SD1 Spare probe head and diaphragm set DOEL SD1 Electrolyte for probe APAPTER ST 9 VDC adapter for 110VAC power supply RS232 SD RS232...

Страница 16: ...15 NOTES...

Страница 17: ...e Jl Radin Inten II No 62 Duren Sawit Jakarta 13440 Indonesia Workshop Jl Pahlawan Revolusi No 22B Jakarta 13430 Indonesia Phone 021 8690 6777 Hunting Fax 021 8690 6777 Mobile 62 816 1740 8925 Fax 021...

Отзывы: