37
8-6 Press
ENT
key to set the settings.
(The screen is changed to the setting screen of reference temperature for accumulation.)
Log mode
The successive accumulation value (unit:
°
Ch) of the
difference between the recorded temperature and reference
temperature for accumulation is displayed. (For details,
see the Fig. on the right.)
However, note that the value recorded last is not included.
There are two kinds of accumulation values, H
accumulation that accumulates the difference higher than
the reference temperature for accumulation and L
accumulation that accumulates the difference lower than
the reference temperature for accumulation.
Move the cursor to a channel and reference temperature for accumulation by the [
] or [
] key.
Then, set the reference temperature by the [<] or [>] key, or the ten-key pad.
The settable temperature may vary depending on the input type.
Input type
setting range of reference temperature for accumulation
K
−
200.0 to 1370.0
°
C
T
−
200.0 to 400.0
°
C
E
−
200.0 to 700.0
°
C
J
−
200.0 to 700.0
°
C
⇒
To return to the previous screen, press [CANCEL] key.
CAUTION
•
The setting is not available if the input signal type is “V”. The setting is available for a channel,
which has not been selected.
•
This setting is not provided in the tag mode.
ログ 番号
各 チャ ンネ ルの入 力種類
各チャ ンネル の積 算基準 温度
Log No.
Reference temperature for
accumulation of each channel
Sensor type of each
channel
温度
積算基準 値
時 間
収録 インター バル
測定値
L積 算値
H積算 値
収録 開始
収録 終了
Temperature
H accumulation value
Measured value
Accumulation
reference value
L accumulation value
Time
Recording start
Recording end