AJ16 User's Guide
• Any traces left by the mixture on the tool must be removed immediately, remove the fiber-optic
handpiece and clean it with a cloth in wipe with alcohol.
• Don’t point the light of the fiber-optic handpiece to the patient's eyes! The emitted light may
cause harm to some patients, such as cataract patients. Generally speaking, the fiber does not
cause lasting damage, but may induce temporary blindness.
• Limited load of dental chair is 135kg, it will not work properly if overload.
• If dental unit work with external equipment to carry out planting work, you should disconnect
the dental chair power supply before working to avoid personal injury due to malfunction and
accidentally touching the control button.
• Only when the high-speed handpiece and low-speed handpiece completely are stopped, the
needle can not be removed, otherwise it will lead to damage to the chuck, it will lead to
personal injury if the needle is out.
• Only use high quality needles and the appropriate size of the bolts.
• Every day, you should check the damage of collect before starting work to determine whether
the needle is firmly in the handpiece.
• After replacing the needle of high-speed handpiece, pull the needle out to confirm that it is
installed in place.
• The diameter of the needle should be between 1.59-1.60mm (ISO1797
Ⅲ
), the maximum
length is 25mm
(
ISO6360-1 Standard
)
.
• Only in the case of the installation of a needle or repair tool, high-speed handpiece can be
used.
• Do not press the button which make the needle off when the tool is in use, the friction between
the button and the air motor impeller will overheat the head, that may cause burnout.
• The organization (tongue, cheek, lips, etc.) of the patient's mouth must be protected by a
suitable method (with a mirror, etc.) to avoid touching the button
• Do not touch the dental light bulb directly, until the lamp cooling, the replacement after wearing
protective gloves, avoid to be burned.
• It is forbidden to touch PC board and manufacturer's electronic components by hand or metal.
• Users should maintain and replace parts which are easy to age (such as handpiece) on time.
• Can not be used in flammable narcotic gas mixed with air or flammable narcotic with oxygen or
nitrous oxide.