47
P/N: 8013093 Rev
:
F
Table 204
The dental unit is intended for use in the electromagnetic environment specified below. The customer
or the user of the dental unit should assure that it is used in such and environment.
Immunity test
GB 9706 test
level
Compliance
level
Electromagnetic environment-guidance
Conducted
BF
GB/T 17626.6
Radiated BF
GB/T 17626.3
3 Vrms
150 kHz
~
80MHz
3V/m
80 MHz
~
2.5 GHz
[V1]V
Portable and mobile RF communications
equipment should be used no closer to any
part no closer to any part of the monitor
including cables, than the recommended
separation distance calculated from the
equation applicable to thefrequency of the
transmitter.
Recommended separation distance
Where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in meters (m). Field strengths from
fixed RF transmitters, as determined by an
electromagnetic site survey, should be less
than the compliance level in each frequency
range. Interference may occur in the vicinity
of equipment marked with the following
symbol:
NOTER 1: At 80 MHz and 800 MHz
,
the higher frequency range applies.
NOTER 2:These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to
fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field
strength in the location in which the equipment is used exceeds the applicable RF compliance level
above, the equipment should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as reorienting or observed measure may be
necessary, such as reorienting or relocating the equipment.
b. Over the frequency range 150 kHz to 80MHz, field strengths should be less than [V1] V/m.
AJ16 User's Guide