Power off:  Press  and  hold  the  power  button  to  enter  the shutdown  interface. 

All the indicator lights are turned off, the detector enters the shutdown countdown, 
and the display is turned off from OFF 004 to OFF 000. After the countdown ends, the 
display  is  turned  off,  and  the  Green  indicator  keep  flashing  for  about  10  seconds, 
then the instrument is turned off.

 

 

 

 

 

 

 

2

.

3 Daily use

 

Factory reset: After the boot process is displayed, the detector enters the facto-

ry  reset  interface.  If  you  need  to  restore  the  factory  settings,  press  the  function 
button twice in rapid succession, otherwise the detector will automatically enter the 
main interface.

 

 

 

 

 

 

Main  interface:  The  main  interface  is  mainly  used  to  display:  gas  name,  gas 

concentration, alarm type, time and battery status. The green indicator flashes at this 
time.

 

Fault indication: When the gas sensor is shorted or broken, the corresponding 

detection channel has a fault indication, as shown in the figure below. At this point, 
the green indicator is off and the yellow fault indicator is keeps on.

 

 

Shutdown interface

 

Factory reset interface

 

Main interface 

Содержание AGH6200

Страница 1: ...1 AIYI Technologies Operator s Manual AGH6200 AGH6200F Version 3 3 1901...

Страница 2: ...tems The design and manufacture of product is strictly follow the international standard and company regulations and each product get a normative QC control to ensure the best quality for you Please r...

Страница 3: ...rior written permission of AIYI Technologies AIYI TECHNOLOGIES MAKES NO WARRANTIES EXPRESS OR IMPLIED INCLUDING WITHOUT LIMITA TION THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICUL...

Страница 4: ...le to the use and maintenance of AGH6200 AGH6200F series gas detector This manual is suitable for the following personnel instrument maintenance engineers field users Attention Before connecting and o...

Страница 5: ...duction 05 1 2 Instrument description 07 1 3 Interface description 08 1 4 Specifications 09 Chapter 2 Operation 2 1 Packing list 10 2 2 Power on off 10 2 3 Daily use 11 2 4 Operation 12 2 5 Charging 1...

Страница 6: ...tor the following are referred to as detectors The main functions and features of the product Detect combustible gas oxygen carbon monoxide and hydrogen sulfide Imported sensors with high sensitivity...

Страница 7: ...ble gas and toxic gas de tection and alarm for petrochemical industry GB 12358 2006 Gas monitors and alarms for workplace General technical requirements GB 16838 2005 Envionmental test methods and sev...

Страница 8: ...light 2 LCD screen 3 Button 4 Calibration kit adapter 5 Sensors 6 Buzzer 7 Alarm LED light 8 Alligator clip 9 Nameplate Function Choose Exit Power ON OFF Menu Confirm Fault Yellow light keeps on Alarm...

Страница 9: ...ected eg Ex O2 CO H2S STEL Short term exposure limit TWA Time weighted average Low alarm High alarm 3 Readings bar Display the current gas concentration 5 Time bar Display current time 4 Unit bar LEL...

Страница 10: ...ol or relative error 10 O2 3 F S H2S 5 mol mol Response time 30s T90 Sampling method Diffusion sampling standard Alarm method Sound light 3 LEDs vibration Operating temperature 40 70 10 55 Operating h...

Страница 11: ...ers the boot pro cess and the full screen flashes 3 times during the boot process and then displays the STEL alarm value TWA alarm value alarm 1 value alarm 2 value and calibration value of each chann...

Страница 12: ...r enters the facto ry reset interface If you need to restore the factory settings press the function button twice in rapid succession otherwise the detector will automatically enter the main interface...

Страница 13: ...hutdown interface as shown in the figure below 2 During the shutdown countdown process press the power and function button at the same time to enter the menu Menu is shown in the figure below 3 Short...

Страница 14: ...calibration interface is shown in the figure below 2 Short press the function button to select the specific channel and the gas name keep flashes 3 Short press the power button to do zero calibration...

Страница 15: ...ort press the power button to do span calibration If the calibration is com pleted the End prompt will be displayed If the calibration fails the Err prompt will be displayed as shown in the figure 4 P...

Страница 16: ...adapter and the ventilation con necting the calibration adapter to the detector tightly and passing the gas of the specified concentration from the vent of the calibration adapter Self test interface...

Страница 17: ...he concentration value will flash for 3 seconds and then the End prompt will be displayed If the test is not normal the Err prompt will flash for 3 seconds and then the gas concentration will be calib...

Страница 18: ...nt the detector from falling or severe vibration It is strictly forbidden to expose the detector to a gas with high concentration corrosive gas etc for a long time otherwise it will cause a serious da...

Страница 19: ...Warranty AIYI Technologies promises that the product will be guaranteed for one year from sale and will be serviced for life If the product fails or any problem during the use please feel free to con...

Страница 20: ...19 AIYI Technologies SCIENTIFIC INNOVATION SPECIALIZED DEDICATED...

Отзывы: