16
AIYI Technologies
2
.
In the main interface, press and hold the function button to enter the bump
test preparation interface. At this time, the self-test indicator flashes, short press
the power button to enter the bump test.
3
.
Short press the function button to select the quick detection zone and the gas
name indicator flashes.
4
.
After a stable readings, press the power button to enter bump test, and the gas
name flashes. About 15 seconds later, if the test is normal, the concentration
value will flash for 3 seconds, and then the "End" prompt will be displayed. If the
test is not normal, the “Err” prompt will flash for 3 seconds, and then the gas
concentration will be calibrated for processing. If the calibration is successful, the
correct concentration value will be displayed for 3 seconds, and then the “End”
prompt will be displayed. If the calibration fails, the “Err” prompt will flash for 3
seconds, then the “End” prompt will be displayed, followed by the sensor fault as
“---”.
5
.
The bump test is end and the detector automatically back to the main interface.
Main interface
Bump test interface
Bump test completed
Bump test failed
Fault indication