GB
Focus™
GB-29
AirSep
®
Corporation
MN172-1 rev. - 03/11
Reserve Oxygen Supply
Your Equipment Provider may recommend another source of supplemental
oxygen therapy in case there is a mechanical failure or a power cut.
In the event of an alarm or you observe that Focus is not working
properly, consult the Troubleshooting section in this manual. If
you cannot resolve the problem, consult your Equipment
Provider.
If you feel discomfort or are experiencing a medical emergency,
seek medical assistance immediately.
Troubleshooting
The Focus product is designed for years of trouble-free use.
If your Focus Portable Oxygen Concentrator fails to operate properly, refer to the
chart on the following pages for possible causes and solutions and, if necessary,
consult your Equipment Provider.
Do not attempt any maintenance other than the possible
solutions listed below.
Problem Probable
Cause
Solution
Alarm condition, Intermittent:
Beep, beep (Amber) light
illuminates simultaneously
and Focus shuts down.
Battery voltage is too
low to operate the
Focus unit.
Connect to DC or an AC
socket immediately.
Alarm condition, Intermittent:
Beep, beep, beep and the
(Amber) alarm light
illuminates simultaneously.
Breathing rate has
exceeded the capacity
of the
Focus
unit.
Reduce activity, and then turn
unit off and back on again to
reset unit. If necessary,
change to another source of
oxygen as available and
contact your Equipment
Provider.