When the service indicator illuminates solid amber, it is time for your
Focus unit to be inspected and/or serviced by your Equipment Provider.
After any necessary service and when the performance is verified by the
Equipment Provider, the service indicator light will be reset.
This unit is not to be used for life support. Geriatric, paediatric,
or any other patients unable to communicate discomfort while
using this unit may require additional monitoring. Patients with
hearing and/or sight impairment(s) may need assistance with
monitoring alarms.
How to Respond to Focus’s Audible Alarm and Indicator Lights
Status Audible
Alarm
Light Indicates Action
Indicator Brief,
continuous
at start–up
(Green) and
(Red) alternate;
then (Green)
continuous light
Focus has been
turned on.
You may begin to
operate your Focus unit.
Indicator No
(Green)
flashes;
on each breath
Focus is delivering
oxygen as a pulse
flow.
Continue using Focus
normally.
Indicator No
(Amber)
continuous
light
Inspection and/or
Service required.
Return unit to Equipment
Provider for inspection
and/or service.
Dual
display
Service
Required
(amber)
Figure 22: Indicator of Required Inspection/Service