![AirClean systems PowderSafe AC730C Скачать руководство пользователя страница 16](http://html.mh-extra.com/html/airclean-systems/powdersafe-ac730c/powdersafe-ac730c_operators-manual_2876281016.webp)
16
Pre-filters
Pre-filters should be checked each month. In most cases, they should be changed every
1 to 3 months. Prefilter replacement is required when discoloration of the prefilter occurs
or when the low airflow alarm comes on. See the “PRE-FILTER CHECK/REPLACEMENT”
section of this manual for complete instructions.
PRE-FILTER CHECK / REPLACEMENT
A. To check the Pre-filter:
1. Look through the sash to the inside of your unit. The pre-filter and pre-filter grid are located at the
back of the unit.
2. Look through the slats on the pre-filter grid. If the prefilter is discolored (the original color is
green), replacement is needed.
B. To replace the Pre-filter:
1. Don the proper PPE for handling the substance that have been used inside the enclosure (refer to
the MSDS sheets).
2. Turn the unit on and make sure that the blower is running.
3. Place an appropriate poly bag on the base of the enclosure. You will place the saturated pre-filter
inside the bag and seal it before removing it from the hood.
MAINTENANCE
Figure 7. AC740C Pre-filter grid.