AC-DS-03 Automated Deactivation System Manual Ver. A03
4
Plumbing Installation
1. Once the unit is secured to wall, proceed with plumbing hook up.
2. Cut the 3/8” drain tubing (supplied) to required length and install a right angle barbed fitting
(supplied) on both ends of the tubing. The drain DISCHARGE outlet is on the left side of the
unit, tubing can be routed under the unit if your drain is on the right side of the AC-DS-03.
3. Install the 3/8” drain tubing between the AC-DS-03 quick disconnect marked “DISCHARGE”
on the left side of the unit and a suitable drain (by pushing firmly into DISCHARGE
connector.) Make certain that the outlet end of the hose is secure so as not to cause an
accidental spill if discharging into a sink. If drain is on the opposite side of unit, route drain-
tube behind or under AC-DS-03.
4. If connecting the AC-DS-03 to an automatic disinfector such as a TD-100 install discharge
line from the AC-DS-03 into the drain fitting supplied with the unit. See Figure 2.
5. Route the feed line from the supplying unit into the connector on the right side of the AC-DS-
03 marked INLET. As above, route the feed line behind or under unit if required.
6. The 3/8” overflow tube can be connected to either side. Use the 3/8’ straight barbed
connectors to attach the overflow tubing to the AC-DS-03 OVERFLOW connector. Make
sure that the drain is lower than the outlets for the OVERFLOW on the AC-DS-03. There are
overflow connectors on both sides of the unit.
CAUTION!
The overflow connectors are for safety in case of pump failure and will
allow tank to drain and not overflow. The overflow connection must be
used.
CAUTION!
INSTALL THE PLUG (SUPPLIED IN THE INSTALLATION KIT) INTO THE
OPPOSITE SIDE OVERFLOW CONNECTOR. Failure to do this can
cause spillage in the case of pump failure. See Figure 3.
7. Make sure outlet of the overflow tube is secured to a proper drain. If connecting to an
automatic disinfector such as a TD-100 route the overflow tube into the top of the drain fitting
supplied. See Figure 2.
8. If using a transfer pump, connect the outlet of the pump to the inlet of the AC-DS-03. Make
certain the overflow is connected to the drain.
9. Make sure all connections are tightly installed. Unit is now ready to run.
Electrical Requirements
CAUTION!
Plug power cord into hospital grade GFCI protected wall receptacle.