AC-DS-03 Automated Deactivation System Manual Ver. A03 

 

Plumbing Installation

 

 
 
1.  Once the unit is secured to wall, proceed with plumbing hook up. 
 
2.  Cut the 3/8” drain tubing (supplied) to required length and install a right angle barbed fitting 

(supplied) on both ends of the tubing. The drain DISCHARGE outlet is on the left side of the 
unit, tubing can be routed under the unit if your drain is on the right side of the AC-DS-03. 

 
3.  Install the 3/8” drain tubing between the AC-DS-03 quick disconnect marked “DISCHARGE” 

on the left side of the unit and a suitable drain (by pushing firmly into DISCHARGE 
connector.)  Make certain that the outlet end of the hose is secure so as not to cause an 
accidental spill if discharging into a sink.  If drain is on the opposite side of unit, route drain-
tube behind or under AC-DS-03. 

 
4.  If connecting the AC-DS-03 to an automatic disinfector such as a TD-100 install discharge 

line from the AC-DS-03 into the drain fitting supplied with the unit.  See Figure 2. 

 
5.  Route the feed line from the supplying unit into the connector on the right side of the AC-DS-

03 marked INLET.  As above, route the feed line behind or under unit if required. 

 

 

6.  The 3/8” overflow tube can be connected to either side.  Use the 3/8’ straight barbed 

connectors to attach the overflow tubing to the AC-DS-03 OVERFLOW connector.  Make 
sure that the drain is lower than the outlets for the OVERFLOW on the AC-DS-03.  There are 
overflow connectors on both sides of the unit. 

 

 

 

CAUTION!  

The overflow connectors are for safety in case of pump failure and will 
allow tank to drain and not overflow.  The overflow connection must be 
used.   

 

 

CAUTION! 

INSTALL THE PLUG (SUPPLIED IN THE INSTALLATION KIT) INTO THE 
OPPOSITE SIDE OVERFLOW CONNECTOR.  Failure to do this can 
cause spillage in the case of pump failure.  See Figure 3. 

 

7.  Make sure outlet of the overflow tube is secured to a proper drain.  If connecting to an 

automatic disinfector such as a TD-100 route the overflow tube into the top of the drain fitting 
supplied.  See Figure 2.

 

 

8.  If using a transfer pump, connect the outlet of the pump to the inlet of the AC-DS-03.  Make 

certain the overflow is connected to the drain. 

 

9.  Make sure all connections are tightly installed.  Unit is now ready to run.

 

 

 
 

 

Electrical Requirements

 

 

 

 

CAUTION! 

Plug power cord into hospital grade GFCI protected wall receptacle. 

 
 

 

 

Содержание AC-DS-03

Страница 1: ......

Страница 2: ...AC DS 03 Automated Deactivation System Instruction Manual ...

Страница 3: ...03 Printed in the United States of America No part of this publication may be reproduced without the express written permission of AirClean Systems AirClean Systems 2179 E Lyon Station Road Creedmoor NC 27522 USA Tel 800 849 0472 919 255 3220 Fax 919 255 6120 Email info aircleansystems com Web www aircleansystems com ...

Страница 4: ...Ratings 1 Warning Symbols 2 Setting up your AC DS 03 2 Securing your AC DS 03 3 Plumbing Installation 4 Electrical Requirements 4 Filter Installation and Maintenance 6 Use of AC DS 03 Automated Deactivation System 6 Cleaning 8 Monthly Cleaning 8 Quick Trouble Shoot 8 Main Wiring Diagram 8 Limited Warranty 9 Specifications and Technical Data 9 ...

Страница 5: ... 03 can be used in conjunction with automatic processing devices for endoscopic ultrasound probes Safety Precautions Disconnect power before replacing or servicing the unit Use PPE whenever operating the AC DS 03 wear safety goggles lab coat proper gloves and face eye shield Environmental Conditions For indoor use only Temperature range 5 C to 40 C Main supply voltage fluctuations not to exceed 10...

Страница 6: ...AL EQUIPMENT DAMAGE Protective Grounding Setting up your AC DS 03 Included with your new AC DS 03 system is a packet of the following parts Power cord Two E Z mount wall anchors and screws Two wall mounting brackets 10 feet of 3 8 O D drain hose 10 feet of 3 8 O D overflow hose Two Right angle 3 8 barbed adaptors Two Straight 3 8 barbed adaptors One port plug One sample packet of NeutraHyde Part A...

Страница 7: ... 03 and install screws loosely 4 Position unit against wall for wall bracket installation Push bracket against wall and mark position of bracket hole See Figure 1 5 Repeat on other side of unit 6 Install the two self drilling wall anchors at these points until they are flush with the wall Place bracket over anchor and install Phillips head screws do not over tighten Now tighten 7 For masonry walls...

Страница 8: ...ght side of the AC DS 03 marked INLET As above route the feed line behind or under unit if required 6 The 3 8 overflow tube can be connected to either side Use the 3 8 straight barbed connectors to attach the overflow tubing to the AC DS 03 OVERFLOW connector Make sure that the drain is lower than the outlets for the OVERFLOW on the AC DS 03 There are overflow connectors on both sides of the unit ...

Страница 9: ...tomated Deactivation System Manual Ver A03 5 Figure 3 3 1 Typical installations of the AC DS 03 PLEASE CALL AIRCLEAN SYSTEMS AT 1 800 849 0472 WITH ANY INSTALLATION QUESTIONS YOU MAY HAVE Inlet Overflow Discharge ...

Страница 10: ...m is designed to be used for neutralizing high level disinfectants before disposal CAUTION Always use proper PPE whenever operating the AC DS 03 1 Once the preceding steps including filter installation have been followed the unit is ready to be plugged in 2 Plug the power cord into the power receptacle on the left side of unit 3 Plug power cord into hospital grade 20 Amp GFCI protected wall recept...

Страница 11: ...ng 11 Once tank is empty the ADD NEUTRALIZER LED will flash and operator will repeat cycle starting at step 5 above 12 If tank needs to be emptied any time before reaching the full level the OPTIONAL DISCHARGE button can be pushed at any time to pump out tank contents 13 Unit will then return to step 5 to initiate cycle Figure 4 LED reset and discharge buttons Figure 5 Access cover on top of unit ...

Страница 12: ... not running check the following 1 Ensure that the unit is plugged into a live wall outlet 2 Make sure the On Off switch located front panel is on 3 In the event that the system is still not working check the two fuses The fuses are located on the left side of the AC DS 03 in a pull out drawer on the power receptacle CAUTION UNPLUG UNIT BEFORE PULLING OPEN FUSE DRAWER 4 Pull open drawer and remove...

Страница 13: ...es obligations or other liabilities expressed or implied What is not covered by this warranty Conditions and damages resulting from any of the following Improper installation delivery within the facility or maintenance Any repair modification alteration or adjustment not authorized by AirClean Systems Misuse abuse accidents or unreasonable use Incorrect electric current voltage or supply Failure t...

Отзывы: