PREPARE THE VACSTAR TO RECEIVE THE HYDROMISER ASSEMBLY
Note: 36” height clearance required for proper vent installation
A. Disconnect the power and shut off the water supply.
B. Disconnect the water distribution system
(See Fig 1)
.
1. Disconnect both 3/8” poly flow tubes at the water distribution system.
2. The other end of the 3/8” poly flow tubes are connected to the pump housing. Measure 13-1/2”
from the tube fittings on the pump housing. Cut off the excess tubing.
3. Disconnect both 1/4” poly flow tubes at the bottom of the water distribution system.
Note: When reconnecting these tubes, they must be reconnected in the same way they were
disconnected as one tube is longer than the other.
4. Remove and discard both water distribution systems from the back rail using a Phillips screw driver
to disengage the clamp.
C. Remove and discard the existing exhaust hose and fitting between the exhaust
manifold and the drain.
Cut off the exhaust hose. Use adjustable pliers to unscrew the rest of the fitting (See Figs. 1 & 6).