Page 9
Air Techniques, Inc.
Air Techniques, Inc.
Page 8
INSTALLATION INFORMATION
Installation Notes.
Installation Layout Space.
Figure 4 shows the requirements for the installation of MOJAVE
LT3 . Please note that all units are shipped with leveling feet set to lowest position . Heights can be
increased by 1 inch by adjusting the leveling feet . Refer to Figures 5 and 6 for connection details .
A. PUMP INSTALLATION SPACE -
Area for
MOJAVE LT3
pump in typical installation .
B. SEWER DRAIN -
Provide a drain for the removal of waste liquids from the Air/Water Separator assembly .
Use an open drain pipe (1 ½ inch P-Trap with 1 inch air gap or floor sink) or a closed vented drain .
See Figure 6 .
C. HEAT EXHAUST -
Refer to Figure above and see Plumbing Requirements for the exhaust vent line required
for
MOJAVE LT3
. Schedule 40 pipe can be used on
MOJAVE LT3.
D. PUMP ELECTRIC SERVICE -
The
MOJAVE LT3
pump is wired directly with a dedicated 220V, 20 AMP, single
phase 50/60 Hz circuit . If Main Circuit panel is not located in equipment room, a disconnect box with approved
ground is needed for each pump . Disconnect boxes should be mounted no more than 3 feet of the installation
center line .
E. OVERHEAD INSTALLATION VACUUM LINE -
See Plumbing Requirements for
MOJAVE LT3
connection .
F.
SUB FLOOR INSTALLATION VACUUM LINE -
See Plumbing Requirements for
MOJAVE LT3
connection .
48"
4"
4"
20"
Sewer Drain
4"
MOJAVE LT3
INSTALL
AREA
60"
1"
12"
24"
36"
E
A
B
C
D
F
Figure 4. MOJAVE LT3 System Floor Plan