![Air Techniques MOJAVE LT3 Скачать руководство пользователя страница 10](http://html1.mh-extra.com/html/air-techniques/mojave-lt3/mojave-lt3_user-and-installation-manual_2875588010.webp)
Page 11
Air Techniques, Inc.
Air Techniques, Inc.
Page 10
MOJAVE LT3 Connection Procedure.
Using industry standard techniques, make the three
connections between the LT3 with supplied components from accessory kits Refer to Figure 5
for the connection diagram and perform the following procedure .
1 .
Treatment Room
Suction Line to Pump Inlet Connection
. Refer to Figure 5, item (1) .
a . Install one connector adapter to the pipe (either overhead or sub floor) from the operatory .
b . Install the connector adapter into the flexible coupling connector on the LT3 .
c . Install the 1
½
" ID, clear hose cut for installation between the operatory suction line and
pump inlet and secure with two no-hub adapters .
2 .
Heat Exhaust Connection
. Refer to Figure 5, item (2) for the location of the vent hose connection
at the pump heat exchanger output and the facility vent line . Make the heat exhaust connection by
performing the following procedure .
a .
Exhaust Vent Assembly Installation
. Refer to Figure 6 and install the Drip Leg and Exhaust
Vent Assembly to the bottom end of the facility vent line . Install a length of 1/4 inch OD
Urethane Tubing (P/N 51453) between the vent condensation drain port and facility sewer
drain .
b .
Heat Exhaust Vent Connection
. As shown by Figure 6, connect the vent hose
between the pump heat exchanger output and the facility vent line . Secure with no hub
adapters provided by the accessory kit .
Important:
Make sure to efficiently use space by making connections as short and direct as possible
to meet your particular site requirements.
Make sure that all hose connections are straight and secure without any sharp bends or kinks.
Since the vacuum hose is rigid, make sure not to stress connections especially at the pump inlet.
INSTALLATION
Heat
Exhaust to
Outside
Vent
Inlet Flexible Coupling
Connector
Liquids/Solids
from the Air/
Water Separator
to Sewer Drain
Figure 5. MOJAVE LT3 Plumbing Connection
1
3
2