User’s and
Installation Manual
Dry Vacuum System
RECOMMEND DAILY
USE OF MONARCH
CLEANSTREAM
Страница 1: ...User s and Installation Manual Dry Vacuum System RECOMMEND DAILY USE OF MONARCH CLEANSTREAM...
Страница 2: ...ing 24 Warranty 26 On Line Warranty Registration 26 TABLE OF CONTENTS Figure Title Page 1 MOJAVE LT3 Key Parts Identification 5 2 MOJAVE LT3 Functional Connection Points 6 3 MOJAVE LT3 Vacuum Pump Dim...
Страница 3: ...conserving electricity and prolonging the life of your pump SAFETY SUMMARY Use of the MOJAVE LT3 not in conformance with the instructions specified in this manual may result in permanent failure of th...
Страница 4: ...s Operating Temperature 50 to 105 F or 10 to 40 C Storage and Transport Temperature 32 to 158 F or 0 to 70 C Operating Relative Humidity 80 no condensation Storage and Transport Relative Humidity 90 n...
Страница 5: ...avels through the exhaust vent The LCD touch screen provides the operational user interface for the MOJAVE LT3 system PRODUCT DESCRIPTION Figure 1 MOJAVE LT3 Key Parts Identification Front View Parts...
Страница 6: ...quency vacuum level RPM reading and temperature as well as recording the run time in hours Automatic Vacuum Adjustment When an instrument suction tip has been opened the system senses an increase in v...
Страница 7: ...ute Drain tube size 1 inch The drain line should be a short run with a maximum rise of 3 feet Avoid any sharp right angle bends Make sure to install the supplied exhaust vent assembly to the bottom en...
Страница 8: ...t Pipe 1 PVC Sch 40 Minimum Suction Line Pipe 1 PVC Sch 40 Maximum Suction Line Pipe See note 2 1 PVC Sch 40 Riser Pipe PVC Sch 40 Vacuum Line Termination 1 Spigot Branch Line Pipe Size requirement of...
Страница 9: ...a closed vented drain See Figure 6 C HEAT EXHAUST Refer to Figure above and see Plumbing Requirements for the exhaust vent line required for MOJAVE LT3 Schedule 40 pipe can be used on MOJAVE LT3 D PUM...
Страница 10: ...connection by performing the following procedure a Exhaust Vent Assembly Installation Refer to Figure 6 and install the Drip Leg and Exhaust Vent Assembly to the bottom end of the facility vent line I...
Страница 11: ...in pipe method 4 System Electrical Connections Refer to the Electrical Connections section See Figure 7 and connect the pump to 220V facility power Refer to Figures 8 and 9 when connecting the low pow...
Страница 12: ...Make the 24 VDC connections shown by Figure 9 View B for Remote Panel Switch 53201 1 When using a switch not provided by Air Techniques all remote system status indication is lost Figure 7 MOJAVE LT3...
Страница 13: ...tch Installation ELECTRICAL CONNECTIONS Figure 9 3 Wire and 4 Wire Remote Switch Connection Note Use 18 Gauge for interconnect cable to connect between unit and remote switch YEL ORN BRN 2 ORN YEL RED...
Страница 14: ...single convenient location within the dental office using an optional Remote Control Panel switch Note The motor power circuit breaker must be kept in the ON position and should not be used as a switc...
Страница 15: ...that the when in Standby with no errors the push button indicator is extinguished Note a Depending on the site installation the remote switch can be either a Bi Color LED 6VDC switch included or a sin...
Страница 16: ...me Screen Standby Button that when blue indicates the system is in the Standby mode When pressed this switch starts the compressor operation Running Button Home Screen Standby Button that turns green...
Страница 17: ...ays current RPM value with min max range indicator c VOLTAGE TEMPERATURE SCREEN i Voltage V Widget Current line voltage with min max range indicator iI Temperature F Widget Current ambient temperature...
Страница 18: ...ARAMETERS i Set Point Sets vacuum level between 8 10 inHg in 0 5 inHg increments by pressing up or down arrows Press the SAVE button to accept setting d DATE TIME i Date Sets to current date Set date...
Страница 19: ...urs ii Run Time Shows time that unit has been running hours g ALARM HISTORY Shows the last forty 40 alarms triggered Push any listed ALARM button to get details of alarm such as suggested tasks and da...
Страница 20: ...are critical problems disabling operation As shown below explanation of the Warning or Error is expanded by pressing the displayed alert Also refer to TROUBLESHOOTING page 24 to correct additional pr...
Страница 21: ...Pressure Sensor Malfunction Alert Example Home Screen Showing Error Example Error Screen b ERRORS TOUCH SCREEN CONTROLS i Pressure sensor malfunction Contact Technician ii Room temperature above 105 F...
Страница 22: ...able equipment operation and reduces breakdown to a minimum it is essential that the maintenance instructions be followed completely The routine cleaning will remove any built up deposits in the pipin...
Страница 23: ...ge Suction Hose Adapter Required Not Supplied Procedure Prepare the cleaning solution and clean the system daily by performing the following steps Note The CleanStream Dispenser can hold a maximum of...
Страница 24: ...dering number and description for accessory components available to maintain the MOJAVE product family Contact an authorized Air Techniques dealer for information ACCESSORIES OPTIONS Description Part...
Страница 25: ...rature in the equipment room is above 40 F In crease heating to the equipment room if necessary Vacuum level value is more than 1 0 inHg over setpoint a Feedback control problem b Control board malfun...
Страница 26: ...other products for which an RMA has been issued may be returned to the customer at the discretion of Air Techniques Inc Any item returned under warranty will be repaired or replaced at our option at...
Страница 27: ...Page 27 Air Techniques Inc NOTES...
Страница 28: ...zer and Lotion Waterline Cleaner Evacuation System Cleaner Imaging Accessories Chemistry Processor Accessories For over 50 years Air Techniques has been a leading innovator and manufacturer of dental...