Important:

 

The Collection Container is not to remain in use for a period longer than 12 months (one 

year) from the date of installation .  Check container sediment level line weekly to ensure 

the effectiveness of waste collection.

Note:

 

The Collection Container of an 

Acadia Plus

 Amalgam Separator may 

fill faster

 than 

expected especially after a new vacuum pump is installed .  A new vacuum pump may 

have the capability to remove built up deposits from the vacuum lines that would inevi-

tably fill the Separator in less time than anticipated . 

Usage Period of Collection Container

The Collection Container is not to remain in use for a period longer than 12 months (one year) from the date of 

installation .  Effectiveness of the Collection Container is reduced after 12 months due to the common growth of organic 

material within the unit .  If the container becomes full of collected solid dental waste and not replaced, it would result 

in improper separation of amalgam and make it non-compliant to an amalgam inspection .  Consequently, it is recom-

mended that the container sediment level, as shown by Figure 8, be checked on a weekly basis to ensure the effective

-

ness of the separator . The following provides a guideline to predict when to replace the collection container based on 

the number of users . 

CONTAINER REPLACEMENT & DISPOSAL

Check weekly and replace 

container when waste level 

is at or above the FILL line .

Replacement and Disposal of the Collection Container

The use period for the Collection Container of an 

Acadia Plus 

Amalgam Separator is 

12 months

 or less based on the 

above table . At the end of the use period, arrangement 

must

 be made for proper removal and disposal of the Collection 

Container and its contents in accordance with all local, state and/or federal laws and regulations for disposal of dental 

waste containing mercury .

A Recycle Kit for the Collection Container (P/N APREC) is available from Air Techniques . The kit provides container 

recycling instructions (Instruction Sheet, P/N APREC-1), a replacement container,  along with pre-paid shipping labels 

for no-hassle, safe disposal/recycling using our hazardous waste recycling partners . 

Air Techniques also offers a Replacement Kit for the Collection Container (P/N AP REP), which includes only a replace

-

ment container for customers who prefer to use their own hazardous waste recycling merchant . 

A Shipping Kit is available as well that only includes the materials needed to recycle a used Collection Container 

(P/N APKIT), that provides all necessary packaging and pre-paid shipping labels for no-hassle, safe disposal/

recycling without a replacement container .

Collection Container Replacement

Number of Users

Use Period

1 to 2

9 to 12 months

3 to 4

6 to 9 months

5 to 10

6 months or less

CAUTION

No vacuum pressure causes Separator liquid spills .

•  The vacuum system 

must be on

 when 

replacing Collection Container .

Figure 8 . Collection Container Waste Fill Levels

CAUTION

Bleach reduces Separator effectiveness .

•  DO NOT

 use Chlorine Bleach or solutions of sodium Hypochlorite 

to clean or disinfect the Separator or vacuum system .

Содержание Acadia Plus

Страница 1: ...INSTRUCTION MANUAL...

Страница 2: ...nce 13 TABLE OF CONTENTS LIST OF ILLUSTRATIONS Figure Title Page 1 Acadia Plus Amalgam Separator Assembly PN APLUS 3 2 Acadia Plus Amalgam Separator Assembly Outline Dimension Drawing 5 3 Acadia Plus...

Страница 3: ...ental suction system s wastewater stream Installed between dental operatories and vacuum pumps the Separator uses continuous separation of solids from liquids and gases through a sedimentation process...

Страница 4: ...material skin cleansers oxygen and skin disinfectants are present This appliance is not to be used in areas where the atmosphere could cause fire Dental Waste Handling Collection Containers collect wa...

Страница 5: ...s shown by label replacement line Maximum vacuum 20 in Hg 677 mbar Classification Type 2 per ISO 11143 Connection type 2 in 5 08 cm no hub 1 in 3 81 cm no hub 1 in 2 54 cm NPT and in 1 91 cm NPT Mount...

Страница 6: ...hat all listed items were received If any item is missing notify Air Techniques Figure 3 Acadia Plus Amalgam Separator Assembly Component Identification Installation Accessory Kit P N A1640 Descriptio...

Страница 7: ...l codes and meet all local state and or federal environmental laws and regulations Figure 4 Typical Acadia Plus Amalgam Separator Installation Connections Acadia Plus Separator Inlet Connection Acadia...

Страница 8: ...ary Note Make sure to save all removed screws They are reused during installation 1 Remove the 4 screws securing the separator assembly to the chassis mounting bracket 2 Rotate the separator assembly...

Страница 9: ...tant The Separator assembly must be removed from the chassis in order to mount the bracket to single wall stud Refer to Figure 5 When mounting directly to a single stud use the two Center Mounting Hol...

Страница 10: ...Installation Refer to Figure 7 and install the Collection Container as follows 1 Align the two tabs of the container with the corresponding slots of the collar 2 Lift the Collection Container into pla...

Страница 11: ...of the Collection Container The use period for the Collection Container of an Acadia Plus Amalgam Separator is 12 months or less based on the above table At the end of the use period arrangement must...

Страница 12: ...6 2 inch ID Clear with Blue Helix Spiral Hose Note If more than 6 feet of hose is needed order additional P N 54512 by the foot 54512 6 2 inch to 1 inch Flexible Coupling H5158 2 inch to 2 inch Flexi...

Страница 13: ...al numbers available Then go to the Air Techniques web site www airtechniques com click the register a product link and complete the registration form This on line registration ensures a record for th...

Страница 14: ...NOTES...

Страница 15: ...NOTES...

Страница 16: ...s Our priority is ensuring complete satisfaction by manufacturing reliable products and providing excellent customer and technical support Whether the need is digital imaging utility room equipment or...

Страница 17: ......

Отзывы: