INSTRUCTION MANUAL

Содержание Acadia Plus

Страница 1: ...INSTRUCTION MANUAL...

Страница 2: ...nce 13 TABLE OF CONTENTS LIST OF ILLUSTRATIONS Figure Title Page 1 Acadia Plus Amalgam Separator Assembly PN APLUS 3 2 Acadia Plus Amalgam Separator Assembly Outline Dimension Drawing 5 3 Acadia Plus...

Страница 3: ...ental suction system s wastewater stream Installed between dental operatories and vacuum pumps the Separator uses continuous separation of solids from liquids and gases through a sedimentation process...

Страница 4: ...material skin cleansers oxygen and skin disinfectants are present This appliance is not to be used in areas where the atmosphere could cause fire Dental Waste Handling Collection Containers collect wa...

Страница 5: ...s shown by label replacement line Maximum vacuum 20 in Hg 677 mbar Classification Type 2 per ISO 11143 Connection type 2 in 5 08 cm no hub 1 in 3 81 cm no hub 1 in 2 54 cm NPT and in 1 91 cm NPT Mount...

Страница 6: ...hat all listed items were received If any item is missing notify Air Techniques Figure 3 Acadia Plus Amalgam Separator Assembly Component Identification Installation Accessory Kit P N A1640 Descriptio...

Страница 7: ...l codes and meet all local state and or federal environmental laws and regulations Figure 4 Typical Acadia Plus Amalgam Separator Installation Connections Acadia Plus Separator Inlet Connection Acadia...

Страница 8: ...ary Note Make sure to save all removed screws They are reused during installation 1 Remove the 4 screws securing the separator assembly to the chassis mounting bracket 2 Rotate the separator assembly...

Страница 9: ...tant The Separator assembly must be removed from the chassis in order to mount the bracket to single wall stud Refer to Figure 5 When mounting directly to a single stud use the two Center Mounting Hol...

Страница 10: ...Installation Refer to Figure 7 and install the Collection Container as follows 1 Align the two tabs of the container with the corresponding slots of the collar 2 Lift the Collection Container into pla...

Страница 11: ...of the Collection Container The use period for the Collection Container of an Acadia Plus Amalgam Separator is 12 months or less based on the above table At the end of the use period arrangement must...

Страница 12: ...6 2 inch ID Clear with Blue Helix Spiral Hose Note If more than 6 feet of hose is needed order additional P N 54512 by the foot 54512 6 2 inch to 1 inch Flexible Coupling H5158 2 inch to 2 inch Flexi...

Страница 13: ...al numbers available Then go to the Air Techniques web site www airtechniques com click the register a product link and complete the registration form This on line registration ensures a record for th...

Страница 14: ...NOTES...

Страница 15: ...NOTES...

Страница 16: ...s Our priority is ensuring complete satisfaction by manufacturing reliable products and providing excellent customer and technical support Whether the need is digital imaging utility room equipment or...