9 - Continuous RF waveform operation
Frequency and amplitude step adjustment
27
TGR2050 Series Instruction Manual
Frequency and amplitude step adjustment
The frequency and amplitude can be live adjusted using a user specified step. The
frequency step size can be set from 10Hz to 1GHz and the amplitude step size from
0.1dB to 140dB/ 0.1uV to 998.8mV.
The amplitude step size can be set in dB or linear units (mV or uV) irrespective of the
current output amplitude setting units. A dB step size can be used to adjust the
output amplitude even when it is set in linear units. Similarly, a step size set in uV or
mV can be used to adjust the output amplitude even when it is set in dBm or dBuV.
To set the step size, first open the utility menu by pressing UTILITY and select
Instrument
.
Select the
Freq. Step Size/ Ampl. Step Size
parameter window, enter the required
step size and touch
OK
.
Return to the home screen by pressing either the FREQ or AMP key on the front
panel.
The set step size is displayed in the information box on the home screen.
Select the frequency or amplitude parameter window on the home screen and use
the
˄ & ˅
arrow keys to implement the user defined step.
NOTE
For the step function to work, the step value and the set frequency or amplitude
units
do not
need match to operate.