45
600W
<NRF>
Set 600W mode on or off, where
<NRF>
has the meaning:
0=Off (400W mode), 1=On (600W mode).
600W?
Returns the setting of the 600W mode.
The response is: 600W
<NR1><RMT>
where
<NR1>
is either 0 or 1, where
0=400W mode or 1=600W mode.
A
<NRF>
Set Level A to
<NRF>
. The units are implied by the present load mode.
B
<NRF>
Set Level B to
<NRF>
. The units are implied by the present load mode.
A?
Return the set Level of Level A.
The response is: A
<NR2>
U
<RMT>
where the
<NR2>
is followed by a units suffix determined by the load mode.
B?
Return the set Level of Level B.
The response is: B
<NR2>
U
<RMT>
where the
<NR2>
is followed by a units suffix determined by the load mode.
DROP
<NRF>
Set the Dropout Voltage level to
<NRF>
, in Volts.
DROP?
Returns the set Dropout Voltage level.
The response is: DROP
<NR2>
V
<RMT>
where
<NR2>
is in Volts.
SLEW
<NRF>
Set the Slew rate to
<NRF>
, in unit/s (units of the present Load Mode) using an
exponent as required.
SLEW?
Returns the set Slew rate.
The response is: SLEW
<NR3>
U
<RMT>
where
<NR3>
is in Unit/s with the Unit
determined by the present load mode and an exponent part as required.
SLOW
<NRF>
Set Slow Start facility on or off, where
<NRF>
has the meaning: 0=Off, 1=On.
SLOW?
Returns the setting of the Slow Start facility.
The response is: SLOW
<NR1><RMT>
where
<NR1>
is either 0 (=Off) or 1 (= On)
LVLSEL
<CPD>
Set the Active Level Select to
<CPD>
Where
<CPD>
can be A, B, T, V or E
corresponding to Level A, Level B, Transient, Ext Voltage and Ext TTL.
LVLSEL?
Returns the Level Select state.
The response is: LVLSEL
<CRD><RMT>
Where
<CRD>
can be A, B, T, V or E
corresponding to Level A, Level B, Transient, Ext Voltage and Ext TTL.
FREQ
<NRF>
Set the Transient Frequency to
<NRF>
, in Hz.
FREQ?
Returns the set Transient Frequency.
The response is: FREQ
<NR2>
HZ
<RMT>
where
<NR2>
is in Hz.
DUTY
<NRF>
Set the Transient Duty Cycle (%A) to
<NRF>
, in percent (rounded to integer).
DUTY?
Returns the set Transient Duty Cycle (%A).
The response is: DUTY
<NR1>
%
<RMT>
where
<NR1>
is a percentage.
VLIM
<NRF>
VLIM
<CPD>
Set the Voltage Limit to
<NRF>
. Either VLIM 0 or alternatively
VLIM NONE disables and removes any voltage limit.
VLIM?
Returns the Voltage limit. The response is:
VLIM
<NR2>
V
<RMT>
where
<NR2>
is in Volts, or
VLIM 0V
<RMT>
if no voltage limit is set.
ILIM
<NRF>
ILIM
<CPD>
Set the Current limit to
<NRF>
. Either ILIM 0 or alternatively
ILIM NONE disables and removes any current limit.
ILIM?
Returns the Current Limit. The response is either:
ILIM
<NR2>
A
<RMT>
where
<NR2>
is in Amps, or
ILIM 0A
<RMT>
if no current limit is set.