53
Vesta Charger Wireless Nominal Specifications
Characteristic
Nominal
MICS MedRadio
Frequency Band
402
–
405 MHz Medical Implant
Communication Service (MICS)
Medical Device Radio Communication
Service (MedRadio)
Bandwidth
< 145 kHz
Modulation
FSK
Radiated Power
< 25 µW E.I.R.P.
Range
0 to at least 1.5 m
Transcutaneous Energy Transfer
Frequency Band
13.56 MHz
Industrial, Scientific, and Medical radio
band (ISM)
Bandwidth
< 0.014 MHz
Modulation
Amplitude (slow to optimize coupling, no
data transmitted)
Radiated Power
< 0.6 W
Range
5 mm to 40 mm