52
APPENDIX II
Wireless Technology
RF wireless technology is used in the communication between an
OPTIMIZER Smart Mini Implantable Pulse Generator (IPG) and
the Vesta Charger. It occurs through an encrypted channel over
an RF link that complies with the requirements of the Medical
Implant Communication System (MICS) (range specified to 2 m,
402
–
405 MHz) of the MedRadio Band.
RF wireless technology is also used to transcutaneously transmit
energy from the Vesta Charger to recharge the OPTIMIZER
Smart Mini IPG at the 13.56 MHz ISM frequency. The
transmission range is specified at a maximum of 1.5 inches
between the C
harger’s coil and the IPG’s receiving coil. Control
over the recharge process, as well as the communications of alert
messages from the IPG to the Charger take place over the
encrypted MICS channel.
OPTIMIZER Smart Mini IPG Wireless Nominal Specifications
Characteristic
Nominal
OPTI
link
MICS MedRadio
Frequency Band
402
–
405 MHz Medical Implant
Communication Service (MICS)
Medical Device Radio Communication
Service (MedRadio)
Bandwidth
< 145 kHz
Modulation
FSK
Radiated Power
< 25 µW E.I.R.P.
Range
0 to at least 1.5 m