37
ISO 14117:2019
Clause 4.6 -
Protection from
static magnetic
fields of flux
density up to 1
mT
Per standard
Device operation
is unaffected per
standard
Maintain 6 inches (15 cm)
distance between
household magnets or items
containing magnets (e.g.
headphones, exercise
equipment containing
magnets, etc.) and implant
ISO 14117:2019
Clause 4.7 -
Protection from
static magnetic
fields of flux
density up to 50
mT
Per standard
Does not exhibit
malfunction which
persists after the
removal from the
field per standard
Seek the advice of your
physician or other qualified
health provider regarding
Magnetic Resonance
(NMR), Magnetic
Resonance Imaging (MRI)
•
Exercise caution in
the vicinity of
equipment that
generates strong
magnetic fields.
•
Do not enter an area
with posted warnings
advising pacemaker
patients (or patients
with other types of
implantable devices)
not to approach.
ISO 14117:2019
Clause 4.8 -
Protection from
AC magnetic field
exposure in the
range of 1 kHz to
140 kHz
Per standard
Does not exhibit
malfunction which
persists after the
removal from the
field per standard
Seek the advice of your
physician or other qualified
health provider regarding
Environmental
Conditions
,
Industrial
Machinery
, and
Home
Appliances
.
•
Exercise caution in
the vicinity of
equipment that
generates strong AC
magnetic fields.
•
Do not enter an area
with posted warnings
advising pacemaker
patients (or patients
with other types of
implantable devices)
not to approach.